EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-SA 4.0

Development and Validation of Stability Indicating RP-HPLC Method for the Simultaneous Estimation of Dapagliflozin Propanediol and Metformin Hydrochloride in Tablet Dosage Form

Nikita Shaileshbhai Patel, Bhavesh Hirabhai Patel

International Journal of Pharmaceutical Sciences and Drug Research · 30.11.2020 · 660-667

DOI: 10.25004/ijpsdr.2020.120609 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Pharmaceutical Sciences and Drug Research, online erschienen am 30.11.2020. Verfasst von Nikita Shaileshbhai Patel und 1 weiteren Autor:innen. Die Arbeit wurde bisher 15-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A rapid, precise, accurate, specific and simple stability indicating RP-HPLC method was developed for simultaneous estimation of Dapagliflozin Propanediol (DAPA) and Metformin Hydrochloride (MET) in its tablet dosage form. Method was performed on a column C8 Thermoquest, hypersil division of dimension 250 mm × 4.60 mm having particle size 5 micron. The mobile phase used in the method was 10 mM Ammonium Acetate buffer (pH- 4), Methanol and Acetonitrile in proportion of 30:65:05 respectively. The drug was subjected to acid and alkali hydrolysis, oxidation, photolysis and heat as stress conditions. The method was validated for specificity, linearity, range, precision, accuracy, robustness, LOD, LOQ and system suitability. The flow rate was maintained at 0.8 ml/ min and effluent was monitored at 227 nm. The retention time were observed 5.988 min and 4.661 min for DAPA and MET respectively. The standard curve was found linear over range of 60-140 μg/ml for DAPA and 300-700 μg/ml for MET with correlation coefficient of 0.9996 for DAPA and 0.9994 for MET. The limit of Detection (LOD) of this method was 1.121 μg/ml for DAPA and 6.162 μg/ml for MET. The limit of Quantitation (LOQ) of this method was 3.396 μg/ml for DAPA and 18.674 μg/ml for MET. The percentage recovery was found to be in the range of 98-102% at three different levels of a standard addition. The precision (repeatability, intra-day and inter-day) of the method was within the limit (RSD

Zitieren

APA

Patel, N. S., & Patel, B. H. (2020). Development and Validation of Stability Indicating RP-HPLC Method for the Simultaneous Estimation of Dapagliflozin Propanediol and Metformin Hydrochloride in Tablet Dosage Form. International Journal of Pharmaceutical Sciences and Drug Research, 660-667. https://doi.org/10.25004/ijpsdr.2020.120609

Vancouver

Patel NS, Patel BH. Development and Validation of Stability Indicating RP-HPLC Method for the Simultaneous Estimation of Dapagliflozin Propanediol and Metformin Hydrochloride in Tablet Dosage Form. International Journal of Pharmaceutical Sciences and Drug Research. 2020:660-667. doi:10.25004/ijpsdr.2020.120609

BibTeX
@article{Patel2020,
  author = {Patel, Nikita Shaileshbhai and Patel, Bhavesh Hirabhai},
  title = {Development and Validation of Stability Indicating RP-HPLC Method for the Simultaneous Estimation of Dapagliflozin Propanediol and Metformin Hydrochloride in Tablet Dosage Form},
  journal = {International Journal of Pharmaceutical Sciences and Drug Research},
  year = {2020},
  pages = {660--667},
  doi = {10.25004/ijpsdr.2020.120609}
}
Kostenlos starten →