Construct Validation of a Low-fidelity Endoscopic Sinus Surgery Simulator
American Journal of Rhinology · 2008-11 · 22 (6) · 642-648
DOI: 10.2500/ajr.2008.22.3245 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift American Journal of Rhinology, online erschienen am 01.11.2008. Verfasst von Randy M Leung und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head and Neck Surgery, University of Toronto, Toronto, Canada. Hauptthemen laut MEDLINE: Computer Simulation. Die Arbeit wurde bisher 36-mal zitiert (Crossref) und nennt 10 Literaturangaben.
Abstract (Originalsprache)
Background
Before a simulator becomes widely accepted, it must be relevant, affordable, and accessible. We have developed a low-cost model emphasizing the basic skills required for endoscopic sinus surgery (ESS). It is noninvasive, free from risk of infection, and an excellent low-pressure learning opportunity. The current study was designed to assess the construct validity of our simulator.
Methods
We conducted a stratified randomized crossover-control study. Otolaryngology residents, fellows, and faculty performed predetermined tasks on the model or cadaver, and then switched. Evaluation included hand motion analysis, task time, and blinded expert review.
Results
Sixteen subjects at various levels of training participated. Cadaver performance correlated well with level of training and previous experience with ESS. However, model performance did not demonstrate statistically significant correlation.
Conclusion
Our model was unable to demonstrate clear construct validity at this time. Materials and structural modifications are in progress. Pending further validation, its low-cost construction possesses potential for integration into otolaryngology residency curricula. Assessment of the simulator's ability to improve surgical skill is also planned.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Otolaryngology-Head and Neck Surgery, University of Toronto, Toronto, Canada.
Zitieren
APA
Leung, R. M., Leung, J., Vescan, A., Dubrowski, A., & Witterick, I. (2008). Construct Validation of a Low-fidelity Endoscopic Sinus Surgery Simulator. American Journal of Rhinology, 22(6), 642-648. https://doi.org/10.2500/ajr.2008.22.3245
Vancouver
Leung RM, Leung J, Vescan A, Dubrowski A, Witterick I. Construct Validation of a Low-fidelity Endoscopic Sinus Surgery Simulator. American Journal of Rhinology. 2008;22(6):642-648. doi:10.2500/ajr.2008.22.3245
BibTeX
@article{Leung2008,
author = {Leung, Randy M and Leung, Jerry and Vescan, Allan and Dubrowski, Adam and Witterick, Ian},
title = {Construct Validation of a Low-fidelity Endoscopic Sinus Surgery Simulator},
journal = {American Journal of Rhinology},
year = {2008},
volume = {22},
number = {6},
pages = {642--648},
doi = {10.2500/ajr.2008.22.3245}
}