Aim
To develop the sensitive and specific HPLC-MWD method for the determination of amitriptyline suitable for use in bioanalytical studies in clinical and forensic toxicology. Materials and methods. The analysis was performed on a “MiLiChrome A-02” microcolumn liquid chromatograph attached to a dual-beam multi-wave UV spectrophotometer under the following conditions: 2×75 mm column ProntoSIL 120-5-C18 AQ, 5 μm; the gradient elution mode with a linear gradient, eluent A (5 % acetonitrile and 95 % perchlorate buffer) to eluent B (100 % acetonitrile) for 40 min, then 100 % eluent B for 3 min; the mobile phase flow rate – 100 μL/min; the thermostat temperature – 40°C; detection was carried out at 8 wavelengths (210, 220, 230, 240, 250, 260, 280, 300 nm).
Results
The retention time of amitriptyline was 22.80 min; the absorbance ratios Sλ/S210 were also set as additional option to increase the reliability of identification. The calibration curve was represented by the dependence of a peak area (S, mm2) against concentration (c, μg/μL) and described by the regression equation у = 0.00514 × x. The method showed linearity in the range of 1.0–100.0 μg/mL. The correlation coefficient was lower than 0.99. The LOD and LOQ values were calculated from the parameters of the calibration curve, they were 0.3 μg/mlL and 0.9 μg/mL, respectively. The accuracy and precision of the method developed were determined at three concentration levels (low, middle, high) within one day (intraday); RSD did not exceed 1.0 % across the full concentration range studied when analyzing model solutions.
Conclusions
The sensitive and specific method for the quantitative determination of amitriptyline using the HPLC-MWD method has been developed. The method meets the requirements set to techniques recommended for use in clinical and forensic toxicology, and it has been confirmed by a number of the validation parameters. The HPLC-MWD method developed can be recommended for implementation into the practice of regional bureaus of forensic medical examination and clinical and diagnostic laboratories of regional drug addiction centers.
Zitieren
APA
Baiurka, S. V., & Karpushyna, S. A. (2026). The identification and quantitative determination of amitriptyline by the HPLC-MWD method. News of Pharmacy, 111(1), 74-79. https://doi.org/10.24959/nphj.26.196
Vancouver
Baiurka SV, Karpushyna SA. The identification and quantitative determination of amitriptyline by the HPLC-MWD method. News of Pharmacy. 2026;111(1):74-79. doi:10.24959/nphj.26.196
BibTeX
@article{Baiurka2026,
author = {Baiurka, S. V. and Karpushyna, S. A.},
title = {The identification and quantitative determination of amitriptyline by the HPLC-MWD method},
journal = {News of Pharmacy},
year = {2026},
volume = {111},
number = {1},
pages = {74--79},
doi = {10.24959/nphj.26.196}
}
EUVIMED