EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection

Masaru Matsumoto, Motofumi Suzuki, Mikihiko Karube, Hiroaki Gotoh, Masaya Sato, Tamaki Kobayashi, Minoru Masuda, Sanae Wakita, Atsuo Kawamoto, Yuka Miura, Nao Tamai, Hiromi Sanada

POCUS Journal · 18.11.2026 · 11 (02) · 60-70

DOI: 10.24908/pocusj.v11i02.20395 ↗

Kurz gefasst

Fachartikel in der Zeitschrift POCUS Journal, online erschienen am 18.11.2026. Verfasst von Masaru Matsumoto und 11 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Artificial intelligence (AI) can detect hydronephrosis and estimate prostate volume on ultrasound, but most systems rely on expert-acquired images from conventional console ultrasound systems. We developed an on-device AI model for handheld point-of-care ultrasound (POCUS) and prospectively evaluated it on nurse-acquired scans for bedside triage/decision support.

Methods

The model was trained using retrospective console images (101 renal and 100 prostate cases) and prospective POCUS development clips (44 renal and 88 prostate patients), then evaluated prospectively on ward nurse-acquired renal and suprapubic POCUS videos after brief instruction. Reference standards were blinded expert adjudication of hydronephrosis and expert planimetry for prostate volume. Primary outcomes were sensitivity/specificity for hydronephrosis detection and screening performance for severe prostate enlargement (≥50 mL).

Results

For hydronephrosis detection, 32 evaluable patients were analyzed (one kidney per patient; prevalence 46.9%). The AI achieved a sensitivity of 93.3% (95% CI 70.2–98.8) and a specificity of 82.4% (59.0–93.8). For severe prostate enlargement (≥50 mL; n = 28), sensitivity and specificity were both 71.4%. Agreement for continuous prostate volume was moderate: ICC(A,1) = 0.789; Bland–Altman bias +5.1 mL, with 95% limits −25.0 to +35.1 mL.

Conclusions

On nurse-acquired handheld POCUS, AI showed high sensitivity for hydronephrosis and moderate agreement for prostate-volume estimation, supporting its feasibility for triage support while requiring further validation.

Zitieren

APA

Matsumoto, M., Suzuki, M., Karube, M., Gotoh, H., Sato, M., Kobayashi, T., Masuda, M., Wakita, S., Kawamoto, A., Miura, Y., Tamai, N., & Sanada, H. (2026). Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection. POCUS Journal, 11(02), 60-70. https://doi.org/10.24908/pocusj.v11i02.20395

Vancouver

Matsumoto M, Suzuki M, Karube M, Gotoh H, Sato M, Kobayashi T, et al. Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection. POCUS Journal. 2026;11(02):60-70. doi:10.24908/pocusj.v11i02.20395

BibTeX
@article{Matsumoto2026,
  author = {Matsumoto, Masaru and Suzuki, Motofumi and Karube, Mikihiko and Gotoh, Hiroaki and Sato, Masaya and Kobayashi, Tamaki and Masuda, Minoru and Wakita, Sanae and Kawamoto, Atsuo and Miura, Yuka and Tamai, Nao and Sanada, Hiromi},
  title = {Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection},
  journal = {POCUS Journal},
  year = {2026},
  volume = {11},
  number = {02},
  pages = {60--70},
  doi = {10.24908/pocusj.v11i02.20395}
}
Kostenlos starten →