Iodine based radiopacity of experimental blood clots for testing of mechanical thrombectomy devices
Radiology and Oncology · 2013
DOI: 10.2478/raon-2013-0001 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiology and Oncology, 2013. Verfasst von Zhong Hua Luo und 9 weiteren Autor:innen. Beteiligte Einrichtung: Dotter Interventional Institute, Oregon Health & Science University, Portland, USA ; Department of Radiology, Xijing Hospital, Fourth Military Medica…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Barium sulfate powder used for radiopacity of experimental blood clots (EBCs) for testing mechanical thrombectomy devices (MTD) has negative effects on EBCs mechanical properties. In vitro and in vivo exploration was performed to determine if the iodine based contrast medium will have less negative effects on the EBCs than barium.Materials and methods. Fresh blood from 2 swine was used to create fibrinogen enhanced and thrombin initiated EBC in tubes. Iodine radiopacity was achieved by mixing the blood with 65% Iohexol or by soaking the EBCs for 2 or 24 hours in Iohexol. The EBCs opacified with barium served as controls. In vitro study: The EBCs were subjected to four tests, manual elongation, catheter injection, radiopacity and contrast wash out tests. In vivo study: The common carotid arteries of 2 swine were embolized by either barium EBC or EBC soaked for 24 hours in Iohexol. The duration of radiopacity of the different EBCs was compared.
Results
The EBCs opacified with Iohexol initially had higher radiopacity than the barium opacified EBCs. However, their opacity rapidly decreased with saline soaking and, particularly, after they were embolized in live animals. The mechanical properties of Iohexol opacified EBCs were inferior to barium opacified EBCs. The Iohexol mixed EBCs were less firm and elastic and half of them fragmented during catheter injection. The Iohexol soaked EBCs exhibited decreased tensile strength and elasticity compared to the barium EBCs.
Conclusions
Compared to barium, iodine based contrast medium does not offer any advantage for opacifying EBCs.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Dotter Interventional Institute, Oregon Health & Science University, Portland, USA ; Department of Radiology, Xijing Hospital, Fourth Military Medical University, Xi'an, People's Republic of China.
Zitieren
APA
Luo, Z. H., Chung, A., Choi, G., Lin, Y. H., Pang, H., Uchida, B. T., Pavcnik, D., Jeromel, M., Keller, F. S., & Rösch, J. (2013). Iodine based radiopacity of experimental blood clots for testing of mechanical thrombectomy devices. Radiology and Oncology. https://doi.org/10.2478/raon-2013-0001
Vancouver
Luo ZH, Chung A, Choi G, Lin YH, Pang H, Uchida BT, et al. Iodine based radiopacity of experimental blood clots for testing of mechanical thrombectomy devices. Radiology and Oncology. 2013. doi:10.2478/raon-2013-0001
BibTeX
@article{Luo2013,
author = {Luo, Zhong Hua and Chung, Alex and Choi, Gibok and Lin, Yih Huie and Pang, Huajin and Uchida, Barry T and Pavcnik, Dusan and Jeromel, Miran and Keller, Frederick S and Rösch, Josef},
title = {Iodine based radiopacity of experimental blood clots for testing of mechanical thrombectomy devices},
journal = {Radiology and Oncology},
year = {2013},
doi = {10.2478/raon-2013-0001}
}
EUVIMED