Calibration of Thermodilatometer for Clay Ceramics and Uncertainty Analysis of RLTE and CLTE

Ján Ondruška, Štefan Csáki, Igor Štubňa, Marek Mánik, Anton Trník

Measurement Science Review · 01.12.2026 · 26 (6) · 307-314

DOI: 10.2478/msr-2026-0037 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Measurement Science Review, online erschienen am 21.11.2026. Verfasst von Ján Ondruška und 4 weiteren Autor:innen. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The relative linear thermal expansion (RLTE) and coefficient of linear thermal expansion (CLTE) of a kaolinite-based clay sample were determined using a horizontal alumina push-rod dilatometer in the range 20 °C – 1100 °C. The dilatometer configuration is described along with potential sources of measurement uncertainty. An uncertainty analysis was performed, accounting for temperature-field homogeneity, specimen positioning, mechanical pressure on the sample, thermocouple ageing and corrosion, temperature-control stability, the linear variable differential transformer (LVDT) constant, and the precision of the LVDT output signal. Uncertainties associated with the used instruments were also considered. The summation of individual uncertainty components yields the expanded uncertainty for the thermodilatometric analysis. The uncertainty of RLTE is temperature-dependent, reaching a maximum of ~5.26 % at 800 °C, where the maximum scatter of RLTE was observed. When only ordinary thermal expansion occurs, the uncertainty of the RLTE is ~1 %, and the uncertainty of the CLTE is ~3 %.

Zitieren

APA

Ondruška, J., Csáki, Š., Štubňa, I., Mánik, M., & Trník, A. (2026). Calibration of Thermodilatometer for Clay Ceramics and Uncertainty Analysis of RLTE and CLTE. Measurement Science Review, 26(6), 307-314. https://doi.org/10.2478/msr-2026-0037

Vancouver

Ondruška J, Csáki Š, Štubňa I, Mánik M, Trník A. Calibration of Thermodilatometer for Clay Ceramics and Uncertainty Analysis of RLTE and CLTE. Measurement Science Review. 2026;26(6):307-314. doi:10.2478/msr-2026-0037

BibTeX
@article{Ondruka2026,
  author = {Ondruška, Ján and Csáki, Štefan and Štubňa, Igor and Mánik, Marek and Trník, Anton},
  title = {Calibration of Thermodilatometer for Clay Ceramics and Uncertainty Analysis of RLTE and CLTE},
  journal = {Measurement Science Review},
  year = {2026},
  volume = {26},
  number = {6},
  pages = {307--314},
  doi = {10.2478/msr-2026-0037}
}
Kostenlos starten →