Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements
Journal of Electrical Bioimpedance · 01.01.2026 · 17 (1) · 178-182
DOI: 10.2478/joeb-2026-0017 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Electrical Bioimpedance, online erschienen am 04.11.2026. Verfasst von Jie Hou und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University of Oslo, Oslo, Norway. Die Arbeit nennt 7 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Phase angles measured at low frequencies with the tetrapolar technique are sometimes scattered or even positive. We used Monte Carlo simulations of lock-in detection to quantify how large phase errors that measurement noise alone can cause. For a signal-to-noise ratio after demodulation (SNR out ) of about 3 or higher, the standard deviation of the phase error was close to 57° divided by SNR out , and 95% of the errors were within about twice this value. At SNR out = 10 the error standard deviation was 6°; at SNR out = 1 it was 64° and 16% of the measurements were wrong by more than 90°. As SNR out decreased towards zero, the phase gradually approached a completely random value. In an illustrative low-frequency tetrapolar measurement, the absolute phase errors were small (standard deviation 0.007° at 1 Hz and 0.04° at 0.1 Hz), but the true tissue phase was even smaller. The error was therefore 1.2 times the true phase at 1 Hz and 37 times at 0.1 Hz, and the measured phase had the wrong sign in 20% and 49% of the measurements, respectively. The sign of a small phase angle is only reliable (wrong in less than 1% of measurements) when the true phase is larger than 2.3 times the error standard deviation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physics, University of Oslo, Oslo, Norway.
- Department of Clinical and Biomedical Engineering, Oslo University Hospital, Oslo, Norway.
Zitieren
APA
Hou, J., Abie, S. M., & Martinsen, Ø. G. (2026). Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements. Journal of Electrical Bioimpedance, 17(1), 178-182. https://doi.org/10.2478/joeb-2026-0017
Vancouver
Hou J, Abie SM, Martinsen ØG. Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements. Journal of Electrical Bioimpedance. 2026;17(1):178-182. doi:10.2478/joeb-2026-0017
BibTeX
@article{Hou2026,
author = {Hou, Jie and Abie, Sisay Mebre and Martinsen, Ørjan G},
title = {Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements},
journal = {Journal of Electrical Bioimpedance},
year = {2026},
volume = {17},
number = {1},
pages = {178--182},
doi = {10.2478/joeb-2026-0017}
}
EUVIMED