An artificial intelligence–enabled digital stethoscope demonstrates moderate murmur detection in dogs but not cats and unreliable arrhythmia classification in both species

Jake H Johnson iD, Joshua A Stern iD, Teresa C DeFrancesco iD, Kursten V Pierce iD

Journal of the American Veterinary Medical Association · 01.12.2026 · 264 (12) · 1-8

DOI: 10.2460/javma.26.05.0353 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the American Veterinary Medical Association, erschienen am 01.12.2026. Verfasst von Jake H Johnson und 3 weiteren Autor:innen. Hauptthemen laut MEDLINE: Stethoscopes, Heart Murmurs, Cat Diseases, Dog Diseases. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To prospectively evaluate the diagnostic performance of an AI-enabled digital stethoscope in detecting cardiac murmurs and arrhythmias compared to fourth-year veterinary students and experienced clinicians.

Methods

Dogs and cats presenting to a university teaching hospital were prospectively enrolled from August 1, 2025, through December 31, 2025. Each animal underwent cardiac auscultation at 4 thoracic sites with the use of an AI-enabled digital stethoscope (Core 500; EKO Health Inc), 6-lead ECG, and echocardiogram, if clinically indicated. Auscultation was performed by a cardiology resident, board-certified cardiologist, and fourth-year veterinary student.

Results

The stethoscope demonstrated a sensitivity of 86.8%, specificity of 56.3%, and positive predictive value of 82.5% for murmur detection in dogs. There was no difference between the agreement of the AI stethoscope or students with a clinician (κ = 0.447). In cats, sensitivity was markedly lower (9.1%), with only 2 of 22 murmurs detected (κ = 0.081). The stethoscope demonstrated high sensitivity for atrial fibrillation (100%), but classified no dog as arrhythmia-free. Murmur grade was the only significant predictor of stethoscope murmur diagnosis, and high-grade murmurs (≥ 3) had significantly greater odds of detection (OR, 15.11).

Conclusions

The AI-enabled stethoscope demonstrated moderate murmur detection performance in dogs, comparable to fourth-year veterinary students, but performed poorly in cats. Arrhythmia classification was unreliable in both species. Clinical Relevance AI-enabled stethoscopes represent an emerging tool in veterinary medicine, but clinical validation is necessary before routine adoption. This study provides prospective performance data across small animals and examiner levels, identifying meaningful limitations regarding interpretation of this technology.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

StethoscopesHeart MurmursCat DiseasesDog DiseasesArrhythmias, CardiacHeart AuscultationArtificial IntelligenceAnimalsDogsCatsSensitivity and SpecificityProspective StudiesFemale

Stichwörter der Autor:innen

artificial intelligenceauscultationcardiologymurmurstethoscope

Autor:innen und Institutionen

ORCID-Profile: Jake H Johnson, Joshua A Stern, Teresa C DeFrancesco, Kursten V Pierce

Zitieren

APA

Johnson, J. H., Stern, J. A., DeFrancesco, T. C., & Pierce, K. V. (2026). An artificial intelligence–enabled digital stethoscope demonstrates moderate murmur detection in dogs but not cats and unreliable arrhythmia classification in both species. Journal of the American Veterinary Medical Association, 264(12), 1-8. https://doi.org/10.2460/javma.26.05.0353

Vancouver

Johnson JH, Stern JA, DeFrancesco TC, Pierce KV. An artificial intelligence–enabled digital stethoscope demonstrates moderate murmur detection in dogs but not cats and unreliable arrhythmia classification in both species. Journal of the American Veterinary Medical Association. 2026;264(12):1-8. doi:10.2460/javma.26.05.0353

BibTeX
@article{Johnson2026,
  author = {Johnson, Jake H and Stern, Joshua A and DeFrancesco, Teresa C and Pierce, Kursten V},
  title = {An artificial intelligence–enabled digital stethoscope demonstrates moderate murmur detection in dogs but not cats and unreliable arrhythmia classification in both species},
  journal = {Journal of the American Veterinary Medical Association},
  year = {2026},
  volume = {264},
  number = {12},
  pages = {1--8},
  doi = {10.2460/javma.26.05.0353}
}
Kostenlos starten →