Fachartikel in der Zeitschrift American Journal of Veterinary Research, 2025. Verfasst von Zoltán Szilágyi und 1 weiteren Autor:innen. Beteiligte Einrichtung: VCA Advanced Veterinary Care Center, Fishers, IN. Hauptthemen laut MEDLINE: Intubation, Intratracheal, Anesthesia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To assess the v-gel Advanced Dog supraglottic airway device (SGAD) safety and efficacy compared to a high-volume, low-pressure endotracheal tube (ETT).

Methods

In a prospective randomized study, 30 client-owned mesocephalic canine patients (American Society of Anesthesiologists physical status I or II, scheduled for elective surgery) were to be enrolled and assigned to the SGAD or ETT group (15 SGAD; 15 ETT) by blocked randomization. Endotracheal tube cuffs were inflated to 25 cm H 2 O with a cuff inflator. The primary outcome was anesthetic circuit pressure decrease (leak) of inspiratory air at escalating anesthetic system pressures, tested at 0, 10, 15, and 20 minutes after airway device placement.

Results

Due to safety concerns, the study was terminated after 6 patients in the SGAD group failed leak pressure testing under 20 cm H 2 O at all time points. Thirteen canines (6 SGAD, 7 ETT) were enrolled from September through October 2023. The risk of leaking for the SGAD group was significantly higher than that of the ETT group at 12 cm H 2 O ( P = .005) and at 16 cm H 2 O ( P = .001). There were no perioperative or postoperative adverse events or significant differences in characteristics between airway device groups except that the risk for not requiring manipulations to obtain and maintain an effective airway for the SGAD group was 67% less (relative risk, 0.33; exact 90% CI, 0.06 to 0.74) than that of the ETT group.

Conclusions

The SGAD, but not ETT, failed leak pressure testing < 20 cm H 2 O. Clinical Relevance The v-gel Advanced Dog SGAD was not safe to deliver inhalant anesthetics due to the device failing leak pressure testing under 20 cm H 2 O.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intubation, IntratrachealAnesthesiaAnimalsDogsMaleFemaleProspective Studies

Stichwörter der Autor:innen

anesthetic circuit pressure leakcanineendotracheal tubesupraglottic airway devicev-gel Advanced Dog

Autor:innen und Institutionen

  • VCA Advanced Veterinary Care Center, Fishers, IN.
  • Department of Pathology and Laboratory Medicine, Indiana University School of Medicine, Indianapolis, IN.

Zitieren

APA

Szilágyi, Z., & Szilágyi, K. L. (2025). The v-gel Advanced Dog is inferior to the endotracheal tube for sealing the airway in healthy canine anesthesia. American Journal of Veterinary Research. https://doi.org/10.2460/ajvr.24.11.0359

Vancouver

Szilágyi Z, Szilágyi KL. The v-gel Advanced Dog is inferior to the endotracheal tube for sealing the airway in healthy canine anesthesia. American Journal of Veterinary Research. 2025. doi:10.2460/ajvr.24.11.0359

BibTeX
@article{Szilgyi2025,
  author = {Szilágyi, Zoltán and Szilágyi, Keely L},
  title = {The v-gel Advanced Dog is inferior to the endotracheal tube for sealing the airway in healthy canine anesthesia},
  journal = {American Journal of Veterinary Research},
  year = {2025},
  doi = {10.2460/ajvr.24.11.0359}
}
Kostenlos starten →