EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Accuracy of a flash glucose monitoring system in healthy dogs during isoflurane anesthesia

Elizabeth A Pollack, Lauren N Cromwell, Qianqian Zhao, Adrianna M Sage iD

American Journal of Veterinary Research · 2025

DOI: 10.2460/ajvr.24.08.0242 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Veterinary Research, 2025. Verfasst von Elizabeth A Pollack und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, WI. Hauptthemen laut MEDLINE: Isoflurane, Blood Glucose, Anesthetics, Inhalation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE To assess the accuracy of a flash glucose monitoring system (FGMS; FreeStyle Libre 2) in healthy dogs during isoflurane general anesthesia. METHODS 6 purpose-bred Beagle dogs underwent 3 anesthetic events, 1 week apart, in a randomized crossover study. 1 of 3 mean arterial blood pressure (MAP) states were induced under isoflurane anesthesia (hypotension [MAP, ≤ 60 mm Hg], normotension [MAP, 80 to 120 mm Hg], and hypertension [MAP, ≥ 120 mm Hg]) for 40 minutes. Interstitial glucose was collected from an FGMS (applied to the dog's neck prior to anesthesia) at 0, 5, 10, 20, 30, and 40 minutes from stabilization within targeted MAP range. Whole-blood samples were collected from a lateral saphenous vein for plasma glucose measurement (blood glucose gold standard; BG GS ) and via marginal lip mucosal puncture for point-of-care glucose measurement (blood glucose point-of-care; BG POC ). Shrout-Fleiss intraclass correlation and Bland-Altman analysis were performed. Accuracy was determined using human International Organization for Standardization criteria, including Parkes consensus error grid analysis. RESULTS Interstitial glucose strongly correlated with BG GS and BG POC (intraclass correlation, 0.962 and 0.953, respectively). Interstitial glucose measured higher than BG GS (median[IQR] difference 9[−2 to 20] g/dL) and lower than BG POC (−27[−37 to −11] g/dL). Interstitial glucose BG GS increased as MAP decreased. 62.1% and 63.3% of interstitial glucose measurements were within ± 15 mg/dL when BG GS ≤ 100 mg/dL and within ± 15% of BG GS when > 100 mg/dL, respectively. However, 100% were within zones A and B of error grid analysis. CONCLUSIONS Despite not meeting International Organization for Standardization standards, the FreeStyle Libre 2 demonstrated clinical accuracy for monitoring glucose concentrations in anesthetized dogs. CLINICAL RELEVANCE The application of an FGMS before anesthesia may be useful for monitoring at-risk patients perioperatively.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

IsofluraneBlood GlucoseAnesthetics, InhalationAnimalsDogsCross-Over StudiesFemaleMaleMonitoring, Physiologic

Stichwörter der Autor:innen

FreeStyle Libreanesthesiacanineinterstitial glucosenoninvasive glucose measurement

Substanzen

IsofluraneBlood GlucoseAnesthetics, Inhalation

Autor:innen und Institutionen

  • Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, WI.
  • Department of Biostatistics and Medical Informatics, University of Wisconsin-Madison, Madison, WI.

ORCID-Profile: Adrianna M Sage

Zitieren

APA

Pollack, E. A., Cromwell, L. N., Zhao, Q., & Sage, A. M. (2025). Accuracy of a flash glucose monitoring system in healthy dogs during isoflurane anesthesia. American Journal of Veterinary Research. https://doi.org/10.2460/ajvr.24.08.0242

Vancouver

Pollack EA, Cromwell LN, Zhao Q, Sage AM. Accuracy of a flash glucose monitoring system in healthy dogs during isoflurane anesthesia. American Journal of Veterinary Research. 2025. doi:10.2460/ajvr.24.08.0242

BibTeX
@article{Pollack2025,
  author = {Pollack, Elizabeth A and Cromwell, Lauren N and Zhao, Qianqian and Sage, Adrianna M},
  title = {Accuracy of a flash glucose monitoring system in healthy dogs during isoflurane anesthesia},
  journal = {American Journal of Veterinary Research},
  year = {2025},
  doi = {10.2460/ajvr.24.08.0242}
}