Effect of the surface geometry of uncemented acetabular cups on primary stability in a model of uncemented total hip replacement in dogs

Valentine Pollet, Eric Viguier, Thibaut Cachon

American Journal of Veterinary Research · 2023

DOI: 10.2460/ajvr.23.03.0066 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Veterinary Research, 2023. Verfasst von Valentine Pollet und 2 weiteren Autor:innen. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Acetabulum. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE To compare the in vitro stability of acetabular cups with peripherally reinforced fixation in a model of uncemented total hip replacement in dogs. SAMPLE 63 polyurethane foam blocks and 3 acetabular implant designs: hemiellipsoidal (Model A) and 2 models with equatorial peripheral fins (Model B with 1 level and Model C with 2 levels of fins). METHODS 2 loading patterns—edge loading and push-out tests—were performed until failure and peak forces were recorded. Implantation behavior was visually assessed and the required seating force was assessed using a force-displacement curve. RESULTS Model B showed significantly lower peak force than Model A in edge loading tests with standardized impaction. In the push-out test, Model A had a greater maximal force than Models B and C, with mean maximal forces of 213.7 N, 139.4 N, and 138.9 N, respectively. In the seating force test, Models B and C required a higher force for 2-mm deep implantation (362.0 N and 361.6 N, respectively) than Model A (194.4 N), and were associated with dorsal tilting of the components. CLINICAL RELEVANCE Our results suggest that cups with a peripheral design (B, C) have less primary stability than hemiellipsoidal cups (A). Furthermore, models with peripheral fins (B, C) appeared to have incomplete seating if a higher force was not used during implantation and, therefore, the risk of malpositioning was increased. These data indicate that hemiellipsoidal cups provide the same or better initial stability and require a lower impaction force.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipHip ProsthesisAcetabulumAnimalsDogsBiomechanical PhenomenaProsthesis Design

Zitieren

APA

Pollet, V., Viguier, E., & Cachon, T. (2023). Effect of the surface geometry of uncemented acetabular cups on primary stability in a model of uncemented total hip replacement in dogs. American Journal of Veterinary Research. https://doi.org/10.2460/ajvr.23.03.0066

Vancouver

Pollet V, Viguier E, Cachon T. Effect of the surface geometry of uncemented acetabular cups on primary stability in a model of uncemented total hip replacement in dogs. American Journal of Veterinary Research. 2023. doi:10.2460/ajvr.23.03.0066

BibTeX
@article{Pollet2023,
  author = {Pollet, Valentine and Viguier, Eric and Cachon, Thibaut},
  title = {Effect of the surface geometry of uncemented acetabular cups on primary stability in a model of uncemented total hip replacement in dogs},
  journal = {American Journal of Veterinary Research},
  year = {2023},
  doi = {10.2460/ajvr.23.03.0066}
}
Kostenlos starten →