Variation in eluate volume during DNA extraction does not affect eluate concentration of plasma-derived DNA.
Blood transfusion · 01.09.2026 · 24 (5) · 366-369
DOI: 10.2450/bloodtransfus.1347 ↗Kurz gefasst
Fachartikel in der Zeitschrift Blood transfusion, erschienen am 01.09.2026. Verfasst von Frederik Banch Clausen und 5 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Blood Genetics, Department of Clinical Immunology, Copenhagen University Hospital, Copenhagen, Denmark. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
We have observed significant variation in the volume of eluate from the process of DNA extraction. Such variation may affect the precision of downstream, quantitative DNA measurements. Therefore, we investigated this eluate variation and assessed the potential correlation between DNA concentration and eluate volume in repetitive testing of pooled human plasma sample material. The mean eluate volume of expected 60 µL eluate was 63.3 µL with a coefficient of variation of 6.6%. However, although the total amount of DNA varied and correlated with eluate volume (Spearman rank correlation: r = 0.67), the DNA concentration remained unchanged (p=0.0991), (Spearman rank correlation: r = 0.35). This finding showed that variation in the 60 µL elution volume from the QIAsymphony instrument does not affect downstream, quantitative DNA measurements.
Schlagwörter
Autor:innen und Institutionen
- Laboratory of Blood Genetics, Department of Clinical Immunology, Copenhagen University Hospital, Copenhagen, Denmark.
- Department of Technology, Faculty of Health and Technology, University College Copenhagen, Copenhagen, Denmark.
Zitieren
APA
Clausen, F. B., Jensen, V. B., Kjar, C. L., Thorsen, S. U., Nielsen, L. K., & Krog, G. R. (2026). Variation in eluate volume during DNA extraction does not affect eluate concentration of plasma-derived DNA. Blood transfusion, 24(5), 366-369. https://doi.org/10.2450/bloodtransfus.1347
Vancouver
Clausen FB, Jensen VB, Kjar CL, Thorsen SU, Nielsen LK, Krog GR. Variation in eluate volume during DNA extraction does not affect eluate concentration of plasma-derived DNA. Blood transfusion. 2026;24(5):366-369. doi:10.2450/bloodtransfus.1347
BibTeX
@article{Clausen2026,
author = {Clausen, Frederik Banch and Jensen, Victoria Bølling and Kjar, Caroline Lai and Thorsen, Steffen Ullitz and Nielsen, Leif Kofoed and Krog, Grethe Risum},
title = {Variation in eluate volume during DNA extraction does not affect eluate concentration of plasma-derived DNA},
journal = {Blood transfusion},
year = {2026},
volume = {24},
number = {5},
pages = {366--369},
doi = {10.2450/bloodtransfus.1347}
}