Evaluating metrological performance of a pressure balance utilizing a high-precision force transducer: an experimental study
Metrology and Measurement Systems · 19.08.2024 · 295-305
DOI: 10.24425/mms.2024.149699 ↗Kurz gefasst
Fachartikel in der Zeitschrift Metrology and Measurement Systems, online erschienen am 19.08.2024. Verfasst von Shaker A. Gelany und 1 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).
Abstract (Originalsprache)
This paper describes the development and evaluation of a force-pressure balance system utilizing a highprecision force transducer. This setup aims to calibrate hydraulic pressure gauges commonly used in industrial applications, with a measurement range of up to 50 MPa. Detailed investigations were conducted to determine the metrological characteristics and calibration coefficients of the new force-pressure balance setup. Comparisons were made between the results obtained from the new setup and those obtained from another reference pressure standard. The findings highlighted a decrease in the accuracy of the pressure balance within the lower pressure range. This decrease in accuracy can be attributed to the hysteresis effect caused by the force transducer utilized in the setup. Additionally, an error was observed in the pressure characteristic of the piston’s performance, particularly in the lower pressure range. These findings indicate the need for further improvements in the force-pressure balance system to enhance accuracy across the entire pressure range.
Zitieren
APA
Gelany, S. A., & Mahmoud, G. M. (2024). Evaluating metrological performance of a pressure balance utilizing a high-precision force transducer: an experimental study. Metrology and Measurement Systems, 295-305. https://doi.org/10.24425/mms.2024.149699
Vancouver
Gelany SA, Mahmoud GM. Evaluating metrological performance of a pressure balance utilizing a high-precision force transducer: an experimental study. Metrology and Measurement Systems. 2024:295-305. doi:10.24425/mms.2024.149699
BibTeX
@article{Gelany2024,
author = {Gelany, Shaker A. and Mahmoud, Gouda M.},
title = {Evaluating metrological performance of a pressure balance utilizing a high-precision force transducer: an experimental study},
journal = {Metrology and Measurement Systems},
year = {2024},
pages = {295--305},
doi = {10.24425/mms.2024.149699}
}