KINEMATIC ANALYSIS OF A TENDON–PULLEY MECHANISM FOR A BIONIC PROSTHETIC FINGER
Ukrainian Journal of Mechanical Engineering and Materials Science · 2026 · 12 (3) · 100-114
DOI: 10.23939/ujmems2026.03.100 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ukrainian Journal of Mechanical Engineering and Materials Science, 2026. Verfasst von Vitalii Zaniuk und 1 weiteren Autor:innen. Die Arbeit nennt 17 Literaturangaben.
Abstract (Originalsprache)
Problem statement. Single-actuator prosthetic fingers use tendon transmissions to coordinate several joints, but their motion depends on the changing geometry of the tendon path. Predicting this motion requires a consistent relationship between actuator rotation, joint angles and fingertip displacement.
Purpose
This study develops and numerically evaluates a planar model of a three-phalanx tendon–pulley mechanism with geometric parameters selected to approximate commercial index-finger prosthesis designs. Methodology. The model combines rigid-link kinematics, an inextensible-tendon constraint and moment equilibrium with torsional return springs. The tendon path is represented by straight common-tangent segments and pulley wrap arcs. Differentiating its total length yields configuration-dependent effective moment arms, which determine the distribution of motion between the joints. The equilibrium equations are solved simultaneously by numerical continuation with respect to the drive-pulley angle.
Findings
The resulting curves show that equal spring stiffnesses do not produce equal joint rotations: the changing moment arms redistribute motion throughout flexion. The middle joint reaches its 90° limit at a drive rotation of approximately 430°, while the proximal and distal joint angles are approximately 70° and 77°. This configuration defines the end of the calculated motion branch. At a constant drive speed of 60°/s, fingertip speed ranges from approximately 14 to 33 mm/s. The maximum calculated tendon tension and ideal drive torque are approximately 2.4 N and 10.6 N·mm. Originality. The contribution of the study is a unified description linking the specified pulley arrangement to joint motion, fingertip trajectory and transmission loads. Practical value. These results provide a basis for assessing the kinematic response and quasistatic drive requirements of the selected mechanism. Scopes of further investigations. Further investigations will examine how changes in pulley placement and initial joint configuration affect the predicted motion.
Zitieren
APA
Zaniuk, V., & Gurskyi, V. (2026). KINEMATIC ANALYSIS OF A TENDON–PULLEY MECHANISM FOR A BIONIC PROSTHETIC FINGER. Ukrainian Journal of Mechanical Engineering and Materials Science, 12(3), 100-114. https://doi.org/10.23939/ujmems2026.03.100
Vancouver
Zaniuk V, Gurskyi V. KINEMATIC ANALYSIS OF A TENDON–PULLEY MECHANISM FOR A BIONIC PROSTHETIC FINGER. Ukrainian Journal of Mechanical Engineering and Materials Science. 2026;12(3):100-114. doi:10.23939/ujmems2026.03.100
BibTeX
@article{Zaniuk2026,
author = {Zaniuk, Vitalii and Gurskyi, Volodymyr},
title = {KINEMATIC ANALYSIS OF A TENDON–PULLEY MECHANISM FOR A BIONIC PROSTHETIC FINGER},
journal = {Ukrainian Journal of Mechanical Engineering and Materials Science},
year = {2026},
volume = {12},
number = {3},
pages = {100--114},
doi = {10.23939/ujmems2026.03.100}
}