Technological Intervention to Improve Alarm Management in Acute Care Telemetry Units

Cora R Lehet, Julie A Lopez, Robert J Frank, Maria Cvach

Biomedical Instrumentation & Technology · 2023

DOI: 10.2345/0899-8205-57.2.67 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Instrumentation & Technology, 2023. Verfasst von Cora R Lehet und 3 weiteren Autor:innen. Hauptthemen laut MEDLINE: Clinical Alarms. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Telemetry monitoring is intended to improve patient safety and reduce harm. However, excessive monitor alarms may have the undesired effect of staff ignoring, silencing, or delaying a response due to alarm fatigue. Outlier patients, or those patients who are responsible for generating the most monitor alarms, contribute to excessive monitor alarms.

Methods

Daily alarm data reports at a large academic medical center indicated that one or two patient outliers generated the most alarms daily. A technological intervention aimed at reminding registered nurses (RNs) to adjust alarm thresholds for patients who triggered excessive alarms was implemented. The notification was sent to the assigned RN's mobile phone when a patient exceeded the unit's seven-day average of alarms per day by greater than 400%.

Results

A reduction in average alarm duration was observed across the four acute care telemetry units (P < 0.001), with an overall decrease of 8.07 seconds in the postintervention versus preintervention period. However, alarm frequency increased significantly (χ23 = 34.83, P < 0.001).

Conclusion

Implementing a technological intervention to notify RNs to adjust alarm parameters may reduce alarm duration. Reducing alarm duration may improve RN telemetry management, alarm fatigue, and awareness. More research is needed to support this conclusion, as well as to determine the cause of the observed increase in alarm frequency.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Clinical AlarmsHumansMonitoring, PhysiologicTelemetryIntensive Care UnitsPatient Safety

Zitieren

APA

Lehet, C. R., Lopez, J. A., Frank, R. J., & Cvach, M. (2023). Technological Intervention to Improve Alarm Management in Acute Care Telemetry Units. Biomedical Instrumentation & Technology. https://doi.org/10.2345/0899-8205-57.2.67

Vancouver

Lehet CR, Lopez JA, Frank RJ, Cvach M. Technological Intervention to Improve Alarm Management in Acute Care Telemetry Units. Biomedical Instrumentation & Technology. 2023. doi:10.2345/0899-8205-57.2.67

BibTeX
@article{Lehet2023,
  author = {Lehet, Cora R and Lopez, Julie A and Frank, Robert J and Cvach, Maria},
  title = {Technological Intervention to Improve Alarm Management in Acute Care Telemetry Units},
  journal = {Biomedical Instrumentation & Technology},
  year = {2023},
  doi = {10.2345/0899-8205-57.2.67}
}
Kostenlos starten →