Thermal Disinfection Validation: The Relationship Between A0 and Microbial Reduction
Biomedical Instrumentation & Technology · 2021
DOI: 10.2345/0890-8205-55.3.85 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Instrumentation & Technology, 2021. Verfasst von Terra A. Kremer und 7 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben.
Abstract (Originalsprache)
Validating a thermal disinfection process for the processing of medical devices using moist heat via direct temperature monitoring is a conservative approach and has been established as the A0 method. Traditional use of disinfection challenge microorganisms and testing techniques, although widely used and applicable for chemical disinfection studies, do not provide as robust a challenge for testing the efficacy of a thermal disinfection process. Considerable research has been established in the literature to demonstrate the relationship between the thermal resistance of microorganisms to inactivation and the A0 method formula. The A0 method, therefore, should be used as the preferred method for validating a thermal disinfection process using moist heat.
Zitieren
APA
Kremer, T. A., McDonnell, G., Mitzel, E., Jain, N., Hubert, H., Roth, K., Labrie, P., & Villella, A. (2021). Thermal Disinfection Validation: The Relationship Between A0 and Microbial Reduction. Biomedical Instrumentation & Technology. https://doi.org/10.2345/0890-8205-55.3.85
Vancouver
Kremer TA, McDonnell G, Mitzel E, Jain N, Hubert H, Roth K, et al. Thermal Disinfection Validation: The Relationship Between A0 and Microbial Reduction. Biomedical Instrumentation & Technology. 2021. doi:10.2345/0890-8205-55.3.85
BibTeX
@article{Kremer2021,
author = {Kremer, Terra A. and McDonnell, Gerald and Mitzel, Emily and Jain, Nupur and Hubert, Henri and Roth, Klaus and Labrie, Philippe and Villella, Alex},
title = {Thermal Disinfection Validation: The Relationship Between A0 and Microbial Reduction},
journal = {Biomedical Instrumentation & Technology},
year = {2021},
doi = {10.2345/0890-8205-55.3.85}
}
EUVIMED