EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieMultizentrische Studie

Safety and Feasibility of a Fully Automated Insulin Delivery System: FCL@Home, a Multicenter Randomized Clinical Trial in Individuals With Type 1 Diabetes

Sicherheit und Durchführbarkeit eines vollautomatisierten Insulinabgabesystems: FCL@Home, eine multizentrische randomisierte klinische Studie bei Personen mit Typ-1-Diabetes

Marcela Moscoso-Vasquez iD, Laya Ekhlaspour, Sue A Brown, Mark D DeBoer, Cari Berget, Erin C Cobry, R Paul Wadwa, Gina Capodanno, Eda Cengiz, Jenise C Wong, Elliott C Pryor, Maria F Villa-Tamayo und 15 weitere

Diabetes Care · 01.03.2026 · 49 (3) · 401-409

DOI: 10.2337/dc25-1526 ↗

Auf einen Blick

Ein vollautomatisiertes Closed-Loop-System ohne Mahlzeitenangaben wurde bei Personen mit Typ-1-Diabetes geprüft. Die mittlere Glukose verbesserte sich gegenüber der üblichen Versorgung.

Fragestellung

Ist das AIDANET-System im fully closed-loop (FCL) ohne Mahlzeitenangaben sicher und durchführbar und verbessert es die Glukosekontrolle gegenüber der üblichen Versorgung?

Methode

Randomisierte Crossover-Studie an 3 US-Zentren mit Erwachsenen, jungen Erwachsenen und Jugendlichen; 5 Tage FCL unter Aufsicht und 7 Tage zu Hause; Vergleich mit üblicher Versorgung.

Ergebnisse

34 Teilnehmende schlossen die Studie ab. Mittlere Sensor-Glukose: 178 vs. 164 mg/dL unter UC vs. FCL; P = 0,0058. Zeit im Zielbereich, im engen Zielbereich und oberhalb der Bereiche verbesserte sich, vor allem nachts.

Schlussfolgerung der Autor:innen

Die FCL-Therapie mit dem AIDANET-System kann die durchschnittliche glykämische Kontrolle signifikant verbessern und macht Mahlzeitenboli überflüssig.

Kernaussagen

  • Mittlere Sensor-Glukose verbessert: 178 vs. 164 mg/dL
  • Zeit im Zielbereich und darüber verbesserte sich, vor allem nachts
  • Unterzuckerungszeiten waren unter FCL nicht schlechter als unter UC

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Diabetes Care, online erschienen am 16.12.2025. Verfasst von Marcela Moscoso-Vasquez und 26 weiteren Autor:innen. Beteiligte Einrichtung: Center for Diabetes Technology, University of Virginia, Charlottesville, VA. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1, Insulin, Insulin Infusion Systems, Hypoglycemic Agents. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: JDRF.

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE Automated insulin delivery (AID) requiring manual premeal boluses are the standard of care for people with type 1 diabetes (T1D). We evaluated the safety and feasibility of the Automated Insulin Delivery as an Adaptive NETwork (AIDANET) system used without meal announcements in both supervised and home settings across a diverse population of people with T1D. RESEARCH DESIGN AND METHODS Adults (age >25 years), young adults (age 18–25 years), and adolescents (age 14–18 years) were enrolled at three sites in the U.S. to compare the AIDANET system used in fully closed-loop (FCL) with usual care (UC) in a randomized crossover study. Participants spent ∼5 days using FCL in a supervised environment followed by ∼7 days of use at home. The prespecified primary outcome was defined as the change in mean sensor glucose (MSG) between the 2nd week of UC and the week of the FCL at-home period. RESULTS Overall, 34 participants (12 adults, 10 young adults, 12 adolescents; mean age 25.4 ± 12.6 years; 62% female; mean HbA1c 8.0 ± 1.1%; 94% hybrid closed loop users during UC) completed the study. MSG improved from 178 to 164 mg/dL (UC vs. FCL P = 0.0058). The percentages of time in range (70–180 mg/dL), time in tight range (70–140 mg/dL), and time above range (>180 mg/dL and >250 mg/dL) also significantly improved, primarily overnight. The percentage of time <70 and <54 mg/dL were significantly noninferior in FCL compared with UC. CONCLUSIONS FCL therapy with the AIDANET system can significantly improve average glycemic control while removing the need for mealtime bolusing for a broad population of people with T1D.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diabetes Mellitus, Type 1InsulinInsulin Infusion SystemsHypoglycemic AgentsHumansAdolescentFemaleMaleAdultYoung AdultCross-Over StudiesBlood GlucoseFeasibility Studies

Substanzen

InsulinHypoglycemic AgentsBlood Glucose

Autor:innen und Institutionen

  • Center for Diabetes Technology, University of Virginia, Charlottesville, VA.
  • Division of Endocrinology, Department of Pediatrics, University of California, San Francisco, San Francisco, CA.
  • Barbara Davis Center for Diabetes, University of Colorado Anschutz Medical Campus, Aurora, CO.

ORCID-Profile: Marcela Moscoso-Vasquez, Gregory P Forlenza

Förderung

  • JDRF

Zitieren

APA

Moscoso-Vasquez, M., Ekhlaspour, L., Brown, S. A., DeBoer, M. D., Berget, C., Cobry, E. C., Wadwa, R. P., Capodanno, G., Cengiz, E., Wong, J. C., Pryor, E. C., Villa-Tamayo, M. F., El Fathi, A., Holmes, V., Davis Prince, S., Escobar, E., Towers, L., Hosseininpour, Y., Narayan, A., …, & FCL@Home Study Group* (2026). Safety and Feasibility of a Fully Automated Insulin Delivery System: FCL@Home, a Multicenter Randomized Clinical Trial in Individuals With Type 1 Diabetes. Diabetes Care, 49(3), 401-409. https://doi.org/10.2337/dc25-1526

Vancouver

Moscoso-Vasquez M, Ekhlaspour L, Brown SA, DeBoer MD, Berget C, Cobry EC, et al. Safety and Feasibility of a Fully Automated Insulin Delivery System: FCL@Home, a Multicenter Randomized Clinical Trial in Individuals With Type 1 Diabetes. Diabetes Care. 2026;49(3):401-409. doi:10.2337/dc25-1526

BibTeX
@article{MoscosoVasquez2026,
  author = {Moscoso-Vasquez, Marcela and Ekhlaspour, Laya and Brown, Sue A and DeBoer, Mark D and Berget, Cari and Cobry, Erin C and Wadwa, R Paul and Capodanno, Gina and Cengiz, Eda and Wong, Jenise C and Pryor, Elliott C and Villa-Tamayo, Maria F and El Fathi, Anas and Holmes, Viola and Davis Prince, Sara and Escobar, Estella and Towers, Lindsey and Hosseininpour, Yasaman and Narayan, Avani and Koravi, Chaitanya L K and Nguyen, Taylor and Fulkerson, David and Barnett, Charlotte L and Oliveri, Mary C and Breton, Marc D and Forlenza, Gregory P and FCL@Home Study Group*, FCL@Home Study Group*},
  title = {Safety and Feasibility of a Fully Automated Insulin Delivery System: FCL@Home, a Multicenter Randomized Clinical Trial in Individuals With Type 1 Diabetes},
  journal = {Diabetes Care},
  year = {2026},
  volume = {49},
  number = {3},
  pages = {401--409},
  doi = {10.2337/dc25-1526}
}
Kostenlos starten →