Performance of the Dexcom G6 Continuous Glucose Monitoring System During Cardiac Surgery Using Hypothermic Extracorporeal Circulation

David Herzig, Martina Vettoretti, Dominik P Guensch, Andreas Melmer, Daniel Schürch, Jonathan Roos, Arna M C Goerg, Gabija Krutkyte, Luca Cecchini, Andrea Facchinetti, Andreas P Vogt, Lia Bally iD

Diabetes Care · 01.04.2023 · 46 (4) · 864-867

DOI: 10.2337/dc22-2260 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diabetes Care, online erschienen am 21.02.2023. Verfasst von David Herzig und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diabetes, Endocrinology, Nutritional Medicine and Metabolism, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerl…. Hauptthemen laut MEDLINE: Diabetes Mellitus, Type 1, Cardiac Surgical Procedures. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 14 Literaturangaben. Gefördert von: Dexcom, Swiss Foundation of Anaesthesiology and Intensive Care, Swiss Helmut Horten Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE Continuous glucose monitoring (CGM) may be challenged by extreme conditions during cardiac surgery using hypothermic extracorporeal circulation (ECC). RESEARCH DESIGN AND METHODS We evaluated the Dexcom G6 sensor in 16 subjects undergoing cardiac surgery with hypothermic ECC, of whom 11 received deep hypothermic circulatory arrest (DHCA). Arterial blood glucose, quantified by the Accu-Chek Inform II meter, served as reference. RESULTS Intrasurgery mean absolute relative difference (MARD) of 256 paired CGM/reference values was 23.8%. MARD was 29.1% during ECC (154 pairs) and 41.6% immediately after DHCA (10 pairs), with a negative bias (signed relative difference: −13.7%, −26.6%, and −41.6%). During surgery, 86.3% pairs were in Clarke error grid zones A or B and 41.0% of sensor readings fulfilled the International Organization for Standardization (ISO) 15197:2013 norm. Postsurgery, MARD was 15.0%. CONCLUSIONS Cardiac surgery using hypothermic ECC challenges the accuracy of the Dexcom G6 CGM although recovery appears to occur thereafter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Diabetes Mellitus, Type 1Cardiac Surgical ProceduresHumansBlood GlucoseBlood Glucose Self-MonitoringReproducibility of Results

Substanzen

Blood Glucose

Autor:innen und Institutionen

  • Department of Diabetes, Endocrinology, Nutritional Medicine and Metabolism, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Department of Information Engineering, University of Padova, Padova, Italy.
  • Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

ORCID-Profile: Lia Bally

Förderung

  • Dexcom
  • Swiss Foundation of Anaesthesiology and Intensive Care
  • Swiss Helmut Horten Foundation

Zitieren

APA

Herzig, D., Vettoretti, M., Guensch, D. P., Melmer, A., Schürch, D., Roos, J., Goerg, A. M. C., Krutkyte, G., Cecchini, L., Facchinetti, A., Vogt, A. P., & Bally, L. (2023). Performance of the Dexcom G6 Continuous Glucose Monitoring System During Cardiac Surgery Using Hypothermic Extracorporeal Circulation. Diabetes Care, 46(4), 864-867. https://doi.org/10.2337/dc22-2260

Vancouver

Herzig D, Vettoretti M, Guensch DP, Melmer A, Schürch D, Roos J, et al. Performance of the Dexcom G6 Continuous Glucose Monitoring System During Cardiac Surgery Using Hypothermic Extracorporeal Circulation. Diabetes Care. 2023;46(4):864-867. doi:10.2337/dc22-2260

BibTeX
@article{Herzig2023,
  author = {Herzig, David and Vettoretti, Martina and Guensch, Dominik P and Melmer, Andreas and Schürch, Daniel and Roos, Jonathan and Goerg, Arna M C and Krutkyte, Gabija and Cecchini, Luca and Facchinetti, Andrea and Vogt, Andreas P and Bally, Lia},
  title = {Performance of the Dexcom G6 Continuous Glucose Monitoring System During Cardiac Surgery Using Hypothermic Extracorporeal Circulation},
  journal = {Diabetes Care},
  year = {2023},
  volume = {46},
  number = {4},
  pages = {864--867},
  doi = {10.2337/dc22-2260}
}
Kostenlos starten →