Comparison of Three Methods of Measuring the Inter-Pupillary Distance
Journal of Ophthalmic and Optometric Sciences · 16.03.2024
DOI: 10.22037/joos.v6i3.43349 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Ophthalmic and Optometric Sciences, erschienen am 16.03.2024. Verfasst von Ali Moradi und 6 weiteren Autor:innen.
Abstract (Originalsprache)
Purpose
The purpose of this study was to compare the results of three methods for measuring the interpupillary distance (IPD), including the pupillary distance (PD) ruler, PD meter, and auto-refractometer.Patients and Methods: This cross-sectional study was conducted among patients admitted to Basir Eye Clinic, Tehran, Iran, between January 2018 and June 2019. Far-distance IPD was measured using the PD ruler, PD meter, and auto-refractometer by determining the midpoint between the pupils. Additionally, near-distance IPD and PD for each eye were measured separately using the PD ruler and PD meter.
Results
A total of 300 patients participated in the study, with 144 being male and 234 being female. The mean far-distance IPD was significantly higher among males (P < 0.001 for all three methods). There was a statistically significant correlation between far IPD and older age (P < 0.001 for all three methods). The mean far-distance IPD measurements using the auto-refractometer, PD meter, and PD ruler were 62.57 ± 3.79 mm, 61.95 ± 3.82 mm, and 61.24 ± 3.76 mm, respectively, indicating a statistically significant difference (P < 0.001). However, no clinically significant difference (1 mm difference for one-sided or 2 mm difference for two-sided measurements) was observed among any of the measurement methods.
Conclusion
Although there were statistically significant differences between the three methods in measuring far-distance IPD, these differences did not reach clinical significance. Therefore, it appears that these three methods can be used interchangeably in a clinical setting.
Schlagwörter
Zitieren
APA
Moradi, A., Haghparast, A., Aghajani, F., Shojaei, A., Khoee, S. H., Noorizadeh, F., & Vafaei, F. (2024). Comparison of Three Methods of Measuring the Inter-Pupillary Distance. Journal of Ophthalmic and Optometric Sciences. https://doi.org/10.22037/joos.v6i3.43349
Vancouver
Moradi A, Haghparast A, Aghajani F, Shojaei A, Khoee SH, Noorizadeh F, et al. Comparison of Three Methods of Measuring the Inter-Pupillary Distance. Journal of Ophthalmic and Optometric Sciences. 2024. doi:10.22037/joos.v6i3.43349