Validation of a Wearable Sensor Prototype for Measuring Heart Rate to Prescribe Physical Activity: Cross-Sectional Exploratory Study

Fernanda Laís Loro iD, Riane Martins iD, Janaína Barcellos Ferreira iD, Cintia Laura Pereira de Araujo iD, Lucio Rene Prade iD, Cristiano Bonato Both iD, Jéferson Campos Nobre Nobre iD, Mariane Borba Monteiro iD, Pedro Dal Lago iD

JMIR Biomedical Engineering · 11.12.2024 · 9 · e57373

DOI: 10.2196/57373 ↗

Kurz gefasst

Querschnittstudie in der Zeitschrift JMIR Biomedical Engineering, online erschienen am 11.12.2024. Verfasst von Fernanda Laís Loro und 8 weiteren Autor:innen. Beteiligte Einrichtung: Graduate Program of Rehabilitation Sciences, Universidade Federal de Ciências da Saúde de Porto Alegre - UFCSPA, Porto Alegre, Brazil. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Wearable sensors are rapidly evolving, particularly in health care, due to their ability to facilitate continuous or on-demand physiological monitoring.

Objective

This study aimed to design and validate a wearable sensor prototype incorporating photoplethysmography (PPG) and long-range wide area network technology for heart rate (HR) measurement during a functional test.

Methods

We conducted a transversal exploratory study involving 20 healthy participants aged between 20 and 30 years without contraindications for physical exercise. Initially, our laboratory developed a pulse wearable sensor prototype for HR monitoring. Following this, the participants were instructed to perform the Incremental Shuttle Walk Test while wearing the Polar H10 HR chest strap sensor (the reference for HR measurement) and the wearable sensor. This test allowed for real-time comparison of HR responses between the 2 devices. Agreement between these measurements was determined using the intraclass correlation coefficient (ICC3.1) and Lin concordance correlation coefficient. The mean absolute percentage error was calculated to evaluate reliability or validity. Cohen d was used to calculate the agreement’s effect size.

Results

The mean differences between the Polar H10 and the wearable sensor during the test were –2.6 (95% CI –3.5 to –1.8) for rest HR, –4.1 (95% CI –5.3 to –3) for maximum HR, –2.4 (95% CI –3.5 to –1.4) for mean test HR, and –2.5 (95% CI –3.6 to –1.5) for mean recovery HR. The mean absolute percentage errors were –3% for rest HR, –2.2% for maximum HR, –1.8% for mean test HR, and –1.6% for recovery HR. Excellent agreement was observed between the Polar H10 and the wearable sensor for rest HR (ICC3.1=0.96), mean test HR (ICC3.1=0.92), and mean recovery HR (ICC3.1=0.96). The agreement for maximum HR (ICC3.1=0.78) was considered good. By the Lin concordance correlation coefficient, the agreement was found to be substantial for rest HR (rc=0.96) and recovery HR (rc=0.96), moderate for mean test HR (rc=0.92), and poor for maximum HR (rc=0.78). The power of agreement between the Polar H10 and the wearable sensor prototype was large for baseline HR (Cohen d=0.97), maximum HR (Cohen d=1.18), and mean recovery HR (Cohen d=0.8) and medium for mean test HR (Cohen d= 0.76).

Conclusions

The pulse-wearable sensor prototype tested in this study proves to be a valid tool for monitoring HR at rest, during functional tests, and during recovery compared with the Polar H10 reference device used in the laboratory setting.

Schlagwörter

Stichwörter der Autor:innen

HRbiosensorheart ratemedical devicemobile phonephysiological monitorwearable devicewearable system

Autor:innen und Institutionen

  • Graduate Program of Rehabilitation Sciences, Universidade Federal de Ciências da Saúde de Porto Alegre - UFCSPA, Porto Alegre, Brazil.
  • Undergraduate Course of Medicine, Universidade Federal de Ciências da Saúde de Porto Alegre, Porto Alegre, Brazil.
  • Department of Physical Therapy, Universidade Federal de Ciências da Saúde de Porto Alegre - UFCSPA, Porto Alegre, Brazil.
  • Graduate Program in Computing Sciences, Universidade do Vale do Rio do Sinos - UNISINOS, Porto Alegre, Brazil.
  • Institute of Informatics, Universidade Federal do Rio Grande do Sul - UFRGS, Porto Alegre, Brazil.

ORCID-Profile: Fernanda Laís Loro, Riane Martins, Janaína Barcellos Ferreira, Cintia Laura Pereira de Araujo, Lucio Rene Prade, Cristiano Bonato Both, Jéferson Campos Nobre Nobre, Mariane Borba Monteiro, Pedro Dal Lago

Zitieren

APA

Loro, F. L., Martins, R., Ferreira, J. B., de Araujo, C. L. P., Prade, L. R., Both, C. B., Nobre, J. C. N., Monteiro, M. B., & Dal Lago, P. (2024). Validation of a Wearable Sensor Prototype for Measuring Heart Rate to Prescribe Physical Activity: Cross-Sectional Exploratory Study. JMIR Biomedical Engineering, 9, e57373. https://doi.org/10.2196/57373

Vancouver

Loro FL, Martins R, Ferreira JB, de Araujo CLP, Prade LR, Both CB, et al. Validation of a Wearable Sensor Prototype for Measuring Heart Rate to Prescribe Physical Activity: Cross-Sectional Exploratory Study. JMIR Biomedical Engineering. 2024;9:e57373. doi:10.2196/57373

BibTeX
@article{Loro2024,
  author = {Loro, Fernanda Laís and Martins, Riane and Ferreira, Janaína Barcellos and de Araujo, Cintia Laura Pereira and Prade, Lucio Rene and Both, Cristiano Bonato and Nobre, Jéferson Campos Nobre and Monteiro, Mariane Borba and Dal Lago, Pedro},
  title = {Validation of a Wearable Sensor Prototype for Measuring Heart Rate to Prescribe Physical Activity: Cross-Sectional Exploratory Study},
  journal = {JMIR Biomedical Engineering},
  year = {2024},
  volume = {9},
  pages = {e57373},
  doi = {10.2196/57373}
}