Workflow for Fluorescence-Targeted Lamella Milling From Vitrified Cells With a Coincident Fluorescence, Electron, and Ion Beam Microscope

Elise G Perton, Daan B Boltje, Arjen J Jakobi, Jacob P Hoogenboom

BIO-PROTOCOL · 2025 · 15 (1376)

DOI: 10.21769/bioprotoc.5390 ↗

Kurz gefasst

Fachartikel in der Zeitschrift BIO-PROTOCOL, 2025. Verfasst von Elise G Perton und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands. Die Arbeit wurde bisher 4-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Cryo-electron tomography (cryo-ET) is the main technique to image the structure of biological macromolecules inside their cellular environment. The samples for cryo-ET must be thinner than 200 nm, which is not compatible with micron-sized cells. A focused ion beam (FIB), in conjunction with a scanning electron microscope (SEM) to navigate the sample, can be used to ablate material from vitrified cells such that a thin lamella remains. However, the preparation of lamellae with a FIB-SEM is blind to the location of specific cellular structures and biomolecules. Furthermore, the thickness and uniformity of lamella, while crucial for high-quality tomograms, cannot be established accurately with the FIB-SEM. These limitations strongly affect the success rate for cryo-ET on FIB-milled lamellae and thereby the total throughput of the workflow. To mitigate these problems, a coincident light, electron, and ion beam cryo-microscope was developed by retrofitting a fluorescence microscope, cryogenic microcooler, and piezo stage on a FIB-SEM. The fluorescence of molecules of interest can be monitored in real time while milling to ensure the final lamella contains the structure of interest. In addition, reflected light microscopy can be used for thickness and quality control of the lamella. In this protocol, we will describe how the coincident microscope can be used to prepare lamellae from vitrified cells. Key features • Step-by-step protocol for fluorescence-guided FIB-milling with a coincident three-beam cryogenic microscope as described in [1]. • Details about sample loading and unloading, as well as the lamella milling workflow with graphical explanations. • Quality control of lamella, including thickness, uniformity, and ice contamination.

Schlagwörter

Stichwörter der Autor:innen

CLEMCoincident microscopeCryo LM-FIB-SEMCryo-ETCryo-SECOMENZEL systemFIB-SEMFluorescence targeted millingLamella thickness controlQuality controlTEM

Autor:innen und Institutionen

  • Department of Imaging Physics, Delft University of Technology, Delft, The Netherlands.
  • Kavli Institute of Nanoscience, Delft University of Technology, Delft, The Netherlands.
  • Delmic B.V., Delft, The Netherlands.

Zitieren

APA

Perton, E. G., Boltje, D. B., Jakobi, A. J., & Hoogenboom, J. P. (2025). Workflow for Fluorescence-Targeted Lamella Milling From Vitrified Cells With a Coincident Fluorescence, Electron, and Ion Beam Microscope. BIO-PROTOCOL, 15(1376). https://doi.org/10.21769/bioprotoc.5390

Vancouver

Perton EG, Boltje DB, Jakobi AJ, Hoogenboom JP. Workflow for Fluorescence-Targeted Lamella Milling From Vitrified Cells With a Coincident Fluorescence, Electron, and Ion Beam Microscope. BIO-PROTOCOL. 2025;15(1376). doi:10.21769/bioprotoc.5390

BibTeX
@article{Perton2025,
  author = {Perton, Elise G and Boltje, Daan B and Jakobi, Arjen J and Hoogenboom, Jacob P},
  title = {Workflow for Fluorescence-Targeted Lamella Milling From Vitrified Cells With a Coincident Fluorescence, Electron, and Ion Beam Microscope},
  journal = {BIO-PROTOCOL},
  year = {2025},
  volume = {15},
  number = {1376},
  doi = {10.21769/bioprotoc.5390}
}
Kostenlos starten →