EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study

Rebecca L. Smith, Simeon J. West, Jason Wilson

The Open Anesthesiology Journal · 31.12.2021 · 15 (1) · 49-58

DOI: 10.2174/2589645802115010049 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Open Anesthesiology Journal, erschienen am 31.12.2021. Verfasst von Rebecca L. Smith und 2 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Peripheral nerve injury during regional anaesthesia may result from accidental intraneural placement of the needle, or forceful needle nerve contact. Intraneural injections are associated with increased resistance to injection, typically >15 psi. The BBraun BSmart™ is an inline mechanical manometer, offering a visual display of injection pressures.

Objective

The primary objective of this study was to determine if using the BBraun BSmart TM manometer successfully prevents 90% of anaesthetists and anaesthetic assistants from injecting at pressures > 15 psi during simulated nerve block.

Methods

This was a prospective observational study involving anaesthetists and anaesthetic assistants. Two 20 ml injections were performed by each participant, once when the BBraun BSmart TM manometer was obscured from view, and once with the manometer visible. A PendoTech PressureMAT TM S recorded injection pressures.

Results

39 participants completed the study, with a total of 78 injections recorded. During the study, 32 peak pressures during the 78 procedures were recorded above the recommended upper limit of 15 psi, 41% of the total injections. The peak pressure rose above 15 psi in 24/39 (62%) injections when the BBraun Bsmart™ manometer was obscured, but only in 8/39 (21%) injections when the manometer was visible.

Conclusion

The BBraun Bsmart™ manometer did not successfully prevent 90% of anaesthetists or anaesthetic assistants from injecting at unsafe pressures. However, using the BBraun BSmart™ did reduce the number of unsafe injection pressures generated by participants. When utilised in conjunction with PNS and ultrasound guidance, this may offer additional safety during peripheral nerve blockade.

Zitieren

APA

Smith, R. L., West, S. J., & Wilson, J. (2021). Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study. The Open Anesthesiology Journal, 15(1), 49-58. https://doi.org/10.2174/2589645802115010049

Vancouver

Smith RL, West SJ, Wilson J. Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study. The Open Anesthesiology Journal. 2021;15(1):49-58. doi:10.2174/2589645802115010049

BibTeX
@article{Smith2021,
  author = {Smith, Rebecca L. and West, Simeon J. and Wilson, Jason},
  title = {Using the BBraun BSmartTM Pressure Manometer to Prevent Unsafe Injection Pressures During Simulated Peripheral Nerve Blockade: A Pilot Study},
  journal = {The Open Anesthesiology Journal},
  year = {2021},
  volume = {15},
  number = {1},
  pages = {49--58},
  doi = {10.2174/2589645802115010049}
}
Kostenlos starten →