EUVIMEDEUVIMED entdecken
Crossref

Is Iterative Reconstruction an Improvement Over Filtered Back Projection in Processing Gated Myocardial Perfusion SPECT?

Kyoung Sook Won, E. Edmund Kim, Martha Mar, Salman Gohar, William Erwin, Wei Wei, Isis Gayed

The Open Medical Imaging Journal · 2008

DOI: 10.2174/1874347100802010017 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Open Medical Imaging Journal, 2008. Verfasst von Kyoung Sook Won und 6 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The impact of iterative reconstruction (IR) on myocardial perfusion imaging (MPI) interpretation and functional results is under investigated.We evaluated the effect of IR on the interpretation and functional results of MPI compared with filtered back-projection (FBP).Material and Methods: Sixty patients with previously acquired MPI were randomly selected.Studies were processed using FBP and 2D ordered-subsets expectation-maximization (OSEM).Two physicians interpreted the images after IR and FBP processing.Results were confirmed by coronary angiography and/or clinical course.Left ventricular ejection fraction (LVEF), end-diastolic volume (EDV) and end-systolic volume (ESV) were calculated with IR and FBP.

Results

The sensitivity, specificity, positive and negative predictive values and accuracy of physician 1 were 95%, 87%, 80%, 97%, and 90% with FBP and 100%, 80%, 72%, 100%, and 87% with IR.For physician 2, they were 95%, 95%, 91%, 97%, and 95% with FBP and 81%, 95%, 90%, 90%, and 90% with IR.There were no statistically significant differences between physicians' interpretations (P=0.71 for FBP and 0.09 for IR).There was good correlation between IR and FBP calculated LVEF (r=0.98),EDV (r=0.99), and ESV (r=0.99) but significant difference per patient (P=0.02,0.03 and 0.02, respectively).

Conclusion

There is no statistically significant difference in interpretation of MPI with IR versus FBP but there were statistically significant differences in functional results.

Zitieren

APA

Won, K. S., Kim, E. E., Mar, M., Gohar, S., Erwin, W., Wei, W., & Gayed, I. (2008). Is Iterative Reconstruction an Improvement Over Filtered Back Projection in Processing Gated Myocardial Perfusion SPECT?. The Open Medical Imaging Journal. https://doi.org/10.2174/1874347100802010017

Vancouver

Won KS, Kim EE, Mar M, Gohar S, Erwin W, Wei W, et al. Is Iterative Reconstruction an Improvement Over Filtered Back Projection in Processing Gated Myocardial Perfusion SPECT?. The Open Medical Imaging Journal. 2008. doi:10.2174/1874347100802010017

BibTeX
@article{Won2008,
  author = {Won, Kyoung Sook and Kim, E. Edmund and Mar, Martha and Gohar, Salman and Erwin, William and Wei, Wei and Gayed, Isis},
  title = {Is Iterative Reconstruction an Improvement Over Filtered Back Projection in Processing Gated Myocardial Perfusion SPECT?},
  journal = {The Open Medical Imaging Journal},
  year = {2008},
  doi = {10.2174/1874347100802010017}
}
Kostenlos starten →