Porównanie jakości obrazowania systemu HyperSight CBCT i konwencjonalnego kV-CBCT na akceleratorze TrueBeam z wykorzystaniem fantomu Catphan
Letters in Oncology Science · 12.06.2026 · 23 (2)
DOI: 10.21641/los.2026.1.1.274 ↗Kurz gefasst
Fachartikel in der Zeitschrift Letters in Oncology Science, online erschienen am 12.06.2026. Verfasst von Karolina Sanocka und 5 weiteren Autor:innen. Die Arbeit nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-ND 4.0).
Abstract (Originalsprache)
The aim of this study was to quantitatively evaluate the imaging quality of the HyperSight CBCT system compared to conventional kV-CBCT, using the Head, Pelvis, and Thorax protocols. Materials and Methods: The Catphan phantom was imaged on two TrueBeam accelerators-equipped with the HyperSight CBCT system and a conventional kV-CBCT system, respectively. High-contrast and low-contrast resolution, image noise, signal-to-noise ratio, uniformity, and consistency of HU values for standard material inserts were assessed.
Results
Compared to conventional kV-CBCT, the HyperSight CBCT system demonstrated lower noise, higher SNR, and better image uniformity in all analyzed protocols. The Head protocol also achieved better high-contrast resolution and improved visibility of low-contrast objects at 1.0% contrast. Analysis of HU values revealed differences in the imaging of individual materials, with conventional kV-CBCT typically achieving higher HU values, particularly for Teflon and Delrin. Low-contrast resolution depended on the protocol used.
Conclusions
HyperSight CBCT improved selected image quality parameters compared to conventional kV-CBCT, particularly noise, SNR, uniformity, and some resolution parameters. The extent of improvement depended on the acquisition protocol and the parameter assessed, confirming the need for separate evaluation of each clinical protocol.
Zitieren
APA
Sanocka, K., Jankowiak, O., Nowak, E., Wróblewicz, Z., Gajny, S., & Pawałowski, B. (2026). Porównanie jakości obrazowania systemu HyperSight CBCT i konwencjonalnego kV-CBCT na akceleratorze TrueBeam z wykorzystaniem fantomu Catphan. Letters in Oncology Science, 23(2). https://doi.org/10.21641/los.2026.1.1.274
Vancouver
Sanocka K, Jankowiak O, Nowak E, Wróblewicz Z, Gajny S, Pawałowski B. Porównanie jakości obrazowania systemu HyperSight CBCT i konwencjonalnego kV-CBCT na akceleratorze TrueBeam z wykorzystaniem fantomu Catphan. Letters in Oncology Science. 2026;23(2). doi:10.21641/los.2026.1.1.274
BibTeX
@article{Sanocka2026,
author = {Sanocka, Karolina and Jankowiak, Olga and Nowak, Ewelina and Wróblewicz, Zuzanna and Gajny, Sebastian and Pawałowski, Bartosz},
title = {Porównanie jakości obrazowania systemu HyperSight CBCT i konwencjonalnego kV-CBCT na akceleratorze TrueBeam z wykorzystaniem fantomu Catphan},
journal = {Letters in Oncology Science},
year = {2026},
volume = {23},
number = {2},
doi = {10.21641/los.2026.1.1.274}
}
EUVIMED