Comparison of IMO vifa24plus(1-2) and Humphrey Field Analyzer 24-2

Yuki Takagi, Ryo Asano, Kanna Yamashita, Yukihiro Sakai, Sho Yokoyama, Kei Ichikawa, Kazuo Ichikawa

Clinical Ophthalmology · 2025-01 · Volume 19 · 301-307

DOI: 10.2147/opth.s506059 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Ophthalmology, 2025-01. Verfasst von Yuki Takagi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Japan Community Healthcare Organization Chukyo Hospital, Nagoya, Aichi, Japan. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to compare the results of the IMO vifa 24plus(1-2) and HFA 24-2 visual field tests. Patients and Methods: We included 52 patients (104 eyes) with glaucoma who visited Chukyo Eye Clinic between June 2023 and March 2024. On the same day, the HFA 24-2 test using the Swedish Interactive Threshold Algorithm Standard and the IMO vifa 24plus(1-2) test using the Ambient Interactive Zippy Estimated strategy were performed. A Wilcoxon signed-rank test was used to compare fixation, false positives, false negatives, and total test time for both eyes between the HFA and IMO vifa tests. Only eyes with reliable test results were selected, and the visual field index (VFI), mean deviation (MD), and pattern standard deviation (PSD) were examined using Spearman's rank correlation coefficient and the Wilcoxon signed-rank test.

Results

The test times for the HFA 24-2 and IMO vifa 24plus(1-2) were 716.83±118.80 and 628.75±142.70 s, respectively, with the IMO vifa being significantly shorter (P

Schlagwörter

Stichwörter der Autor:innen

HFA 24-2IMO vifa 24plus(1-2)glaucomavisual field test

Autor:innen und Institutionen

  • Department of Ophthalmology, Japan Community Healthcare Organization Chukyo Hospital, Nagoya, Aichi, Japan.
  • Chukyo Eye Clinic, Nagoya, Aichi, Japan.
  • Asano Eye Clinic, Nagoya, Aichi, Japan.

Zitieren

APA

Takagi, Y., Asano, R., Yamashita, K., Sakai, Y., Yokoyama, S., Ichikawa, K., & Ichikawa, K. (2025). Comparison of IMO vifa24plus(1-2) and Humphrey Field Analyzer 24-2. Clinical Ophthalmology, Volume 19, 301-307. https://doi.org/10.2147/opth.s506059

Vancouver

Takagi Y, Asano R, Yamashita K, Sakai Y, Yokoyama S, Ichikawa K, et al. Comparison of IMO vifa24plus(1-2) and Humphrey Field Analyzer 24-2. Clinical Ophthalmology. 2025;Volume 19:301-307. doi:10.2147/opth.s506059

BibTeX
@article{Takagi2025,
  author = {Takagi, Yuki and Asano, Ryo and Yamashita, Kanna and Sakai, Yukihiro and Yokoyama, Sho and Ichikawa, Kei and Ichikawa, Kazuo},
  title = {Comparison of IMO vifa24plus(1-2) and Humphrey Field Analyzer 24-2},
  journal = {Clinical Ophthalmology},
  year = {2025},
  volume = {Volume 19},
  pages = {301--307},
  doi = {10.2147/opth.s506059}
}
Kostenlos starten →