EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 3.0

Mechanical Improvement of Gas Monitoring System in Monoplace Hyperbaric Chamber to Advance the Safety and Efficacy

Hee Young Lee iD, Yoonsuk Lee, Hyun Kim iD, Jin Hui Paik

Medical Devices: Evidence and Research · 2024-06 · Volume 17 · 217-227

DOI: 10.2147/mder.s465022 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Devices: Evidence and Research, 2024-06. Verfasst von Hee Young Lee und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Emergency Medicine, Yonsei University Wonju College of Medicine, Wonju, Gangwon State, Republic of Korea. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

A Monoplace hyperbaric chamber delivers oxygen to the patient's tissues through breathing. Gas monitoring inside the chamber is important because oxygen (O2) is consumed, and carbon dioxide (CO2) is increased because treatment is performed in a closed volume. This study aimed to advance the safety and efficacy of the monoplace hyperbaric chamber (MHC) through mechanical improvement in a gas monitoring system (GMS).

Methods

First, as the oxygen supply method was changed to the direction of the patient's face, it was compared the values of O2, CO2, humidity, and temperature were measured in the MHC and the GMS when operating at 2.0 atmosphere absolute (ATA) and 3.0 ATA. Second, to evaluate the effects of variables across measuring time, it was analyzed in a 3-way repeated measure ANOVA (10 min.×20 min.×30 min.). Lastly, the values before and after the optimization of the MHC were compared by applying a cooler to prevent temperature rise inside the MHC.

Results

In 2.0 ATA, the average humidity was higher in the MHC than in the GMS (p

Schlagwörter

Stichwörter der Autor:innen

gas monitoring systemmechanical improvementmonoplace hyperbaric chambersafety and efficacy

Autor:innen und Institutionen

  • Department of Emergency Medicine, Yonsei University Wonju College of Medicine, Wonju, Gangwon State, Republic of Korea.
  • Hyperbaric Oxygen Therapy Center, Wonju Severance Christian Hospital, Wonju, Gangwon State, Republic of Korea.
  • Department of Emergency Medicine, Inha University College of Medicine, Incheon, Republic of Korea.

ORCID-Profile: Hee Young Lee, Hyun Kim

Zitieren

APA

Lee, H. Y., Lee, Y., Kim, H., & Paik, J. H. (2024). Mechanical Improvement of Gas Monitoring System in Monoplace Hyperbaric Chamber to Advance the Safety and Efficacy. Medical Devices: Evidence and Research, Volume 17, 217-227. https://doi.org/10.2147/mder.s465022

Vancouver

Lee HY, Lee Y, Kim H, Paik JH. Mechanical Improvement of Gas Monitoring System in Monoplace Hyperbaric Chamber to Advance the Safety and Efficacy. Medical Devices: Evidence and Research. 2024;Volume 17:217-227. doi:10.2147/mder.s465022

BibTeX
@article{Lee2024,
  author = {Lee, Hee Young and Lee, Yoonsuk and Kim, Hyun and Paik, Jin Hui},
  title = {Mechanical Improvement of Gas Monitoring System in Monoplace Hyperbaric Chamber to Advance the Safety and Efficacy},
  journal = {Medical Devices: Evidence and Research},
  year = {2024},
  volume = {Volume 17},
  pages = {217--227},
  doi = {10.2147/mder.s465022}
}
Kostenlos starten →