EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

In Silico Modeling Validation and Contact Pressure Distribution Comparison Analysis of Conventional and Robotic-Assisted Unicompartmental Knee Arthroplasty

Yeseop Park iD, Katherine Elbert iD, Jason Koh iD, Farid Amirouche iD

JBJS Open Access · 2025-10 · 10 (4)

DOI: 10.2106/jbjs.oa.25.00176 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JBJS Open Access, 2025-10. Verfasst von Yeseop Park und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, University of Illinois at Chicago, Chicago, Illinois. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Smith and Nephew Orthopaedics.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Unicompartmental knee arthroplasty (UKA) improves recovery, but optimizing load remains challenging. This study measures tibiofemoral contact pressures after conventional and robotic UKA in cadavers and validates a finite element (FE) model against experimental data.

Methods

Sixteen fresh-frozen cadaveric lower limbs underwent medial UKA with conventional cutting guides (n = 8) or CORI robotic assistance (n = 8). Medial (implanted) and lateral (native) compartment pressures were recorded from 0° to 90° using pressure film sensors. The FE model replicated and was validated against these measurements.

Results

Medial compartment pressure decreased significantly from 1.864 MPa at 0° to 0.252 MPa at 90° (p < 0.05). The lateral compartment showed a similar significant decrease from 0.733 MPa to 0.320 MPa (p < 0.05). No significant differences were observed between conventional and robotic-assisted techniques (p > 0.05). The FE model demonstrated strong agreement with the measured data, with r 2 values of 0.9994 (medial) and 0.9962 (lateral).

Conclusion

Conventional and robotic-assisted UKA techniques demonstrated similar contact pressure profiles. However, increased force and area in the native lateral compartment may predispose to postoperative degeneration. The FE model reliably predicted contact behavior and may be especially useful in refining conventional UKA techniques. Level of Evidence: Therapeutic Level II . See Instructions for Authors for a complete description of levels of evidence.

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, University of Illinois at Chicago, Chicago, Illinois.
  • University of Illinois College of Medicine, Rockford, Illinois.
  • Department of Orthopaedic Surgery, Institute of Orthopaedics and Spine, Endeavor Health (Former Northshore University HealthSystem), an Affiliate of the University of Chicago, Pritzker School of Medicine, Skokie, Illino…

ORCID-Profile: Yeseop Park, Katherine Elbert, Jason Koh, Farid Amirouche

Förderung

  • Smith and Nephew Orthopaedics

Zitieren

APA

Park, Y., Elbert, K., Koh, J., & Amirouche, F. (2025). In Silico Modeling Validation and Contact Pressure Distribution Comparison Analysis of Conventional and Robotic-Assisted Unicompartmental Knee Arthroplasty. JBJS Open Access, 10(4). https://doi.org/10.2106/jbjs.oa.25.00176

Vancouver

Park Y, Elbert K, Koh J, Amirouche F. In Silico Modeling Validation and Contact Pressure Distribution Comparison Analysis of Conventional and Robotic-Assisted Unicompartmental Knee Arthroplasty. JBJS Open Access. 2025;10(4). doi:10.2106/jbjs.oa.25.00176

BibTeX
@article{Park2025,
  author = {Park, Yeseop and Elbert, Katherine and Koh, Jason and Amirouche, Farid},
  title = {In Silico Modeling Validation and Contact Pressure Distribution Comparison Analysis of Conventional and Robotic-Assisted Unicompartmental Knee Arthroplasty},
  journal = {JBJS Open Access},
  year = {2025},
  volume = {10},
  number = {4},
  doi = {10.2106/jbjs.oa.25.00176}
}