EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieMultizentrische StudieOpen Access · CC BY 4.0

Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion

Benjamin D Elder iD, Gregory Mundis iD, Robert Eastlack, Paolo Arrigoni iD, Joerg Franke iD, Robert Lee, David Polly iD, the SILVIA Investigator Group

Journal of Bone and Joint Surgery · 04.11.2026 · 108 (21) · 1703-1711

DOI: 10.2106/jbjs.25.01415 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Journal of Bone and Joint Surgery, online erschienen am 16.07.2026. Verfasst von Benjamin D Elder und 7 weiteren Autor:innen. Beteiligte Einrichtung: Mayo Clinic, Rochester, Minnesota. Hauptthemen laut MEDLINE: Spinal Fusion, Bone Screws. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The purpose of this study was to determine whether placement of triangular titanium implants (TTI) suprajacent to S2-alar-iliac (S2AI) pelvic fixation screws during multilevel spinal fusion surgery for adult spine deformity reduces pelvic fixation failures.

Methods

This prospective, multicenter, international, partially blinded randomized clinical trial included 222 patients scheduled for multilevel spinal fusion surgery with S2AI-based pelvic fixation. In the S2AI+TTI group, the mean age was 64.9 years, 54.1% were female, 3.7% were Hispanic, 5.5% were African American, and 1.8% were Asian. In the S2AI-alone group, the mean age was 64.8 years, 58.4% were female, 3.5% were Hispanic, 7.1% were African American, and 0.9% were Asian. Subjects were randomized to pelvic fixation using S2AI screws alone versus S2AI+TTI. Scheduled follow-up included a 24-month full-spine radiograph and a high-resolution computed tomography (CT) scan, which was independently interpreted.

Results

The primary composite radiographic end point (S2AI screw breakage, lucency, or pullout, or posterior spinal rod breakage distal to S1) occurred less frequently in the S2AI+TTI group than in the S2AI-alone group (11.0% versus 20.4%, p = 0.0415). Except for rod-screw dissociation and rod fracture distal to S1, all components of the primary composite end point occurred less frequently in the S2AI+TTI group. TTI placement did not affect sagittal or coronal spinal alignment parameters. The rate of rod fracture throughout the spinal construct was slightly higher in the S2AI+TTI group (9.2% versus 5.3%) but the difference was not significant (p = 0.197).

Conclusions

Placement of TTI suprajacent to S2AI screws reduced the rate of pelvic fixation failure in patients undergoing multilevel spinal fusion with pelvic fixation. Level of Evidence: Therapeutic Level I . See Instructions for Authors for a complete description of levels of evidence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Spinal FusionBone ScrewsHumansFemaleMiddle AgedTitaniumMaleProspective StudiesAged

Substanzen

Titanium

Autor:innen und Institutionen

  • Mayo Clinic, Rochester, Minnesota.
  • San Diego Spine Foundation, San Diego, California.
  • ASST-Ovest Milanese, Legnano, Lombardy, Italy.
  • Klinikum Magdeburg, Magdeburg, Germany.
  • Royal National Orthopaedic Hospital, Stanmore, United Kingdom.
  • University of Minnesota, Minneapolis, Minnesota.

ORCID-Profile: Benjamin D Elder, Gregory Mundis, Paolo Arrigoni, Joerg Franke, David Polly

Zitieren

APA

Elder, B. D., Mundis, G., Eastlack, R., Arrigoni, P., Franke, J., Lee, R., Polly, D., & the SILVIA Investigator Group (2026). Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion. Journal of Bone and Joint Surgery, 108(21), 1703-1711. https://doi.org/10.2106/jbjs.25.01415

Vancouver

Elder BD, Mundis G, Eastlack R, Arrigoni P, Franke J, Lee R, et al. Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion. Journal of Bone and Joint Surgery. 2026;108(21):1703-1711. doi:10.2106/jbjs.25.01415

BibTeX
@article{Elder2026,
  author = {Elder, Benjamin D and Mundis, Gregory and Eastlack, Robert and Arrigoni, Paolo and Franke, Joerg and Lee, Robert and Polly, David and the SILVIA Investigator Group, the SILVIA Investigator Group},
  title = {Placement of Triangular Titanium Implants Suprajacent to Pelvic Fixation Screws Reduces Pelvic Fixation Failure in Multilevel Spinal Fusion},
  journal = {Journal of Bone and Joint Surgery},
  year = {2026},
  volume = {108},
  number = {21},
  pages = {1703--1711},
  doi = {10.2106/jbjs.25.01415}
}
Kostenlos starten →