EUVIMEDEUVIMED entdecken
Crossref

Sagittal Alignment in Total Knee Arthroplasty

Hao-Ming An iD, Wang Gu iD, Shao-Kui Nan iD, Zheng Liu iD, Rui Li iD, Wei Chai iD

Journal of Bone and Joint Surgery · 06.09.2023 · 105 (17) · 1338-1343

DOI: 10.2106/jbjs.23.00076 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Bone and Joint Surgery, online erschienen am 06.07.2023. Verfasst von Hao-Ming An und 5 weiteren Autor:innen. Beteiligte Einrichtung: Senior Department of Orthopedics, the Fourth Medical Center of PLA General Hospital, Beijing, People's Republic of China. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Knee Prosthesis, Osteoarthritis, Knee. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Sagittal alignment determines the extension and flexion of knee prostheses in total knee arthroplasty (TKA). The definition of the sagittal axes may be different between the Mako TKA system (Stryker) and the conventional manual intramedullary approach. Whether there is any discrepancy between the 2 approaches has not been well studied.

Methods

We retrospectively analyzed 60 full-length computed tomographic (CT) scans of the lower extremities of 54 patients. The femur and tibia were modeled by using Mimics (Materialise). The Mako mechanical axes were determined according to the Mako TKA Surgical Guide. The manual intramedullary axes were determined according to the central axis of the tibial proximal and femoral distal medullary cavities. The femoral, tibial, and combined angular discrepancies were measured in the sagittal plane.

Results

On the femoral side, the Mako mechanical axis was more likely to be located in an extended position relative to the manual intramedullary axis (56 of 60 knees). The median angular discrepancy was 2.46° (interquartile range [IQR], 1.56° to 3.43° [range, −1.06° to 5.24°]). On the tibial side, the Mako mechanical axis was likely to be located in a flexed position relative to the manual intramedullary axis (57 of 60 knees). The median angular discrepancy was 2.40° (IQR, 1.87° to 2.84° [range, −0.79° to 4.20°]). The angular discrepancy of the femoral-tibial sagittal angle was 4.63° (IQR, 3.71° to 5.64° [range, 1.20° to 9.02°]).

Conclusions

Compared with manual TKA, the Mako system is more likely to result in a decreased posterior tibial slope and extension of the femoral prosthesis. It may also influence the evaluation of lower-extremity extension and flexion. When using the Mako system, special attention should be given to these discrepancies. Level of Evidence: Therapeutic Level IV . See Instructions for Authors for a complete description of levels of evidence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeRobotic Surgical ProceduresKnee ProsthesisOsteoarthritis, KneeHumansRetrospective StudiesTibiaFemurKnee Joint

Autor:innen und Institutionen

  • Senior Department of Orthopedics, the Fourth Medical Center of PLA General Hospital, Beijing, People's Republic of China.
  • School of Medicine, Nankai University, Tianjin, People's Republic of China.
  • Chinese PLA Medical School, Beijing, People's Republic of China.
  • National Clinical Research Center for Orthopedics, Sports Medicine and Rehabilitation, Beijing, People's Republic of China.

ORCID-Profile: Hao-Ming An, Wang Gu, Shao-Kui Nan, Zheng Liu, Rui Li, Wei Chai

Zitieren

APA

An, H. M., Gu, W., Nan, S. K., Liu, Z., Li, R., & Chai, W. (2023). Sagittal Alignment in Total Knee Arthroplasty. Journal of Bone and Joint Surgery, 105(17), 1338-1343. https://doi.org/10.2106/jbjs.23.00076

Vancouver

An HM, Gu W, Nan SK, Liu Z, Li R, Chai W. Sagittal Alignment in Total Knee Arthroplasty. Journal of Bone and Joint Surgery. 2023;105(17):1338-1343. doi:10.2106/jbjs.23.00076

BibTeX
@article{An2023,
  author = {An, Hao-Ming and Gu, Wang and Nan, Shao-Kui and Liu, Zheng and Li, Rui and Chai, Wei},
  title = {Sagittal Alignment in Total Knee Arthroplasty},
  journal = {Journal of Bone and Joint Surgery},
  year = {2023},
  volume = {105},
  number = {17},
  pages = {1338--1343},
  doi = {10.2106/jbjs.23.00076}
}