Evaluation of parietal pleural adhesion and invasion in subpleural lung cancer: value of B-mode ultrasound and contrast-enhanced ultrasound
Quantitative Imaging in Medicine and Surgery · 2024-05 · 14 (5) · 3302-3311
DOI: 10.21037/qims-23-1542 ↗Kurz gefasst
Fachartikel in der Zeitschrift Quantitative Imaging in Medicine and Surgery, 2024-05. Verfasst von Yuxin Zhang und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ultrasound, First Affiliated Hospital of Guangzhou Medical University, Guangzhou, China. Die Arbeit wurde bisher 5-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
The parietal pleural adhesion/invasion of lung cancer can contribute substantially to poor prognosis and difficulty in surgery. The value of ultrasound in evaluating the parietal pleural adhesion or invasion (pleural adhesion/invasion) of lung cancer remains uncertain. This study investigated the value of B-mode ultrasound and contrast-enhanced ultrasound (CEUS) in diagnosing parietal pleural adhesion/invasion of subpleural lung cancer.
Methods
The study animals included 40 male New Zealand white rabbits. A rabbit subpleural lung cancer model was constructed by injecting VX2 tumor tissue under ultrasound guidance. In the 1-3 weeks after subpleural lesion formation, parietal pleural adhesion/invasion of the largest subpleural lesion was evaluated with B-mode ultrasound and CEUS by two sonographers. The parietal pleural adhesion/invasion was also determined using the gold standard method of findings from anatomical and pathological examination.
Results
Ultimately, 34 rabbits were subjected to complete ultrasonic evaluation. There were 20 and 14 cases with and without parietal pleural adhesion/invasion, respectively, as confirmed by anatomical and pathological evaluations. The diagnostic sensitivity, specificity, and accuracy of sonographer 1 using B-mode ultrasound were 50.0% [95% confidence interval (CI): 26.0-74.0%], 100%, and 70.6% (95% CI: 54.5-86.7%), respectively; for CEUS, they were 90.0% (95% CI: 75.6-100.0%), 100.0%, and 94.1% (95% CI: 85.8-100.0%), respectively. The diagnostic sensitivity, specificity, and accuracy of sonographer 2 using B-mode ultrasound were 45.0% (95% CI: 21.1-68.9%), 92.9% (95% CI: 77.5-100.0%), and 64.7% (95% CI: 47.8-81.6%), respectively; for CEUS, they were 85.0% (95% CI: 67.9-100.0%), 100.0%, and 91.2% (95% CI: 81.1-100.0%), respectively. The diagnostic accuracy of sonographer 1 was higher with CEUS than with B-mode ultrasound, but not significantly so (94.1% vs. 70.6%; P=0.08). The diagnostic accuracy of sonographer 2 was significantly higher with CEUS than with B-mode ultrasound (91.2% vs. 64.7%; P=0.03). The interrater reliability was higher for CEUS than for B-mode ultrasound (κ=0.941 vs. κ =0.717).
Conclusions
Based on an animal model, B-mode ultrasound and CEUS both exhibited good diagnostic efficacy and interrater reliability in evaluating parietal pleural adhesion/invasion of subpleural lung cancer although CEUS outperformed B-mode ultrasound for both measures.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ultrasound, First Affiliated Hospital of Guangzhou Medical University, Guangzhou, China.
- The State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Disease, the First Affiliated Hospital, Guangzhou Medical University, Guangzhou, China.
- Department of Medicine Ultrasonics, Nanfang Hospital, Southern Medical University, Guangzhou, China.
Zitieren
APA
Zhang, Y., Zhang, Z., Liao, H., Li, M., Xu, C., Liang, Z., He, L., Zhang, S., & Tang, Q. (2024). Evaluation of parietal pleural adhesion and invasion in subpleural lung cancer: value of B-mode ultrasound and contrast-enhanced ultrasound. Quantitative Imaging in Medicine and Surgery, 14(5), 3302-3311. https://doi.org/10.21037/qims-23-1542
Vancouver
Zhang Y, Zhang Z, Liao H, Li M, Xu C, Liang Z, et al. Evaluation of parietal pleural adhesion and invasion in subpleural lung cancer: value of B-mode ultrasound and contrast-enhanced ultrasound. Quantitative Imaging in Medicine and Surgery. 2024;14(5):3302-3311. doi:10.21037/qims-23-1542
BibTeX
@article{Zhang2024,
author = {Zhang, Yuxin and Zhang, Zhanwei and Liao, Haixing and Li, Maohan and Xu, Cong and Liang, Zechun and He, Liantu and Zhang, Shiyu and Tang, Qing},
title = {Evaluation of parietal pleural adhesion and invasion in subpleural lung cancer: value of B-mode ultrasound and contrast-enhanced ultrasound},
journal = {Quantitative Imaging in Medicine and Surgery},
year = {2024},
volume = {14},
number = {5},
pages = {3302--3311},
doi = {10.21037/qims-23-1542}
}
EUVIMED