Stability of Codman Hakim programmable valve pressure settings during 0.23 T portable magnetic resonance imaging: a prospective observational study
Quantitative Imaging in Medicine and Surgery · 2026-02 · 16 (2) · 110-110
DOI: 10.21037/qims-2025-673 ↗Kurz gefasst
Fachartikel in der Zeitschrift Quantitative Imaging in Medicine and Surgery, 2026-02. Verfasst von Long Zeng und 10 weiteren Autor:innen. Beteiligte Einrichtung: Neurosurgery Center, Department of Functional Neurosurgery, The National Key Clinical Specialty, The Engineering Technology Research Center of Educat…. Die Arbeit wurde bisher 1-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Patients with hydrocephalus require multiple cranial imaging sessions for visualization of the ventricles following the implantation of a ventriculoperitoneal shunt (VPS) valve. However, conventional imaging methods have certain limitations: high-field-strength magnetic resonance imaging (MRI) equipment can interfere with the pressure setting of a Codman Hakim programmable valve, while mobile computed tomography (CT) exposes patients and medical staff to radiation. Low-field-strength portable MRI (pMRI) offers a safer and more convenient method for the monitoring of patients with hydrocephalus after VPS surgery. This study aimed to evaluate the stability of valve pressure setting during 0.23 T pMRI and to assess the image quality in patients with hydrocephalus.
Methods
This prospective observational study involved 20 patients with hydrocephalus admitted to Zhujiang Hospital of Southern Medical University between September 2022 and September 2023. Eligible patients had hydrocephalus confirmed by recent routine imaging (1.5 T/3 T MRI or CT) and had undergone implantation of a non-MR-resistant Codman Hakim programmable valve. Participants were then subjected to pMRI (0.23 T), before and after which the valve pressure settings were verified via X-ray. Pressure stability was analyzed with the Wilcoxon signed-rank test. Image quality was evaluated through comparison of the Evans index (EI) measured on pMRI with that on standard CT (performed within 24 hours) via intraclass correlation coefficients (ICCs) and Bland-Altman analysis.
Results
In the in vitro pretest, no significant pressure changes were observed (P=0.161). In the clinical study involving 40 pMRI sessions, the valve pressure setting remained unchanged in 30 (75.0%) sessions. Pressure deviations occurred in 10 (25.0%) sessions, but no change exceeded 20 mmH2O. There was no statistically significant difference between pre- and post-pMRI pressure settings (P=0.5552). Regarding image quality (n=22 paired scans), pMRI showed excellent agreement with CT for EI assessment, with an ICC of 0.981 [95% confidence interval (CI): 0.955-0.992; P
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Neurosurgery Center, Department of Functional Neurosurgery, The National Key Clinical Specialty, The Engineering Technology Research Center of Education Ministry of China on Diagnosis and Treatment of Cerebrovascular Di…
- School of Medicine, Dentistry and Biomedical Sciences, Queen's University Belfast, Belfast, UK.
Zitieren
APA
Zeng, L., Xie, M., Zhu, H., Chen, H., Mofatteh, M., Xiao, Q., Wu, Z., Zhang, S., Xie, T., Xu, Y., & Guo, Y. (2026). Stability of Codman Hakim programmable valve pressure settings during 0.23 T portable magnetic resonance imaging: a prospective observational study. Quantitative Imaging in Medicine and Surgery, 16(2), 110-110. https://doi.org/10.21037/qims-2025-673
Vancouver
Zeng L, Xie M, Zhu H, Chen H, Mofatteh M, Xiao Q, et al. Stability of Codman Hakim programmable valve pressure settings during 0.23 T portable magnetic resonance imaging: a prospective observational study. Quantitative Imaging in Medicine and Surgery. 2026;16(2):110-110. doi:10.21037/qims-2025-673
BibTeX
@article{Zeng2026,
author = {Zeng, Long and Xie, Minghan and Zhu, Hongwei and Chen, Houminji and Mofatteh, Mohammed and Xiao, Qingshan and Wu, Zelin and Zhang, Shizhong and Xie, Tao and Xu, Yuting and Guo, Yanwu},
title = {Stability of Codman Hakim programmable valve pressure settings during 0.23 T portable magnetic resonance imaging: a prospective observational study},
journal = {Quantitative Imaging in Medicine and Surgery},
year = {2026},
volume = {16},
number = {2},
pages = {110--110},
doi = {10.21037/qims-2025-673}
}
EUVIMED