Background

Vertebral compression fractures (VCFs) are among the most common spinal fractures and can lead to chronic pain, impaired mobility, functional disability, and diminished quality of life. This study evaluated the feasibility, early clinical outcomes, and short-term safety of the VCFix Spinal System, a novel bilateral transpedicular stabilization implant for the treatment of VCFs, with or without cement augmentation.

Methods

This retrospective analysis included 20 patients treated with the VCFix Spinal System. Feasibility was evaluated using procedural parameters. Clinical outcomes were assessed using the Numerical Pain Rating Scale (NPRS) and Oswestry Disability Index (ODI) at baseline, postoperatively, and at 1-, 3-, and 6-month follow-up. Quality of life was assessed using the EuroQol 5-Dimension 5-Level (EQ-5D-5L) at baseline and at 1-, 3-, and 6-month follow-up. Postoperative imaging was qualitatively evaluated, and safety outcomes were assessed through 6 months.

Results

All procedures were completed successfully. Mean procedure duration was 25.1±6.5 and 20.0±4.6 minutes for VCFix with and without cement, respectively. The mean NPRS score decreased significantly from 8.5±0.9 at baseline to 1.2±1.0 postoperatively, to 0.6±0.8 at 6 months. Similarly, the mean ODI score decreased significantly from 82.7%±9.5% at baseline to 11.9%±12.9% at postoperatively, to 4.7%±4.8% at 6 months. EQ-5D-5L score improved significantly from 19.8±13.4 at baseline to 92.6±8.5 at 6 months. Non-symptomatic cement leakage was observed in 15% of patients receiving VCFix with cement. Only 10% of patients developed adjacent vertebral fractures. No further procedure- or device-related complications were identified.

Conclusions

The VCFix Spinal System was feasible to implant and demonstrated a favorable early safety profile, with improvements in pain, function, and quality of life observed in patients with VCFs. Larger comparative studies with longer follow-up are recommended to confirm these findings.

Zitieren

APA

Jacobs, Z., Sajadi, B., Ardeshiri, A., & Pflugmacher, R. (2026). Vertebral stabilization using the VCFix Spinal System for vertebral compression fractures: early clinical outcomes. Journal of Spine Surgery, 12(11), 175-175. https://doi.org/10.21037/jss-2026-0240

Vancouver

Jacobs Z, Sajadi B, Ardeshiri A, Pflugmacher R. Vertebral stabilization using the VCFix Spinal System for vertebral compression fractures: early clinical outcomes. Journal of Spine Surgery. 2026;12(11):175-175. doi:10.21037/jss-2026-0240

BibTeX
@article{Jacobs2026,
  author = {Jacobs, Zaina and Sajadi, Banafsheh and Ardeshiri, Ardeshir and Pflugmacher, Robert},
  title = {Vertebral stabilization using the VCFix Spinal System for vertebral compression fractures: early clinical outcomes},
  journal = {Journal of Spine Surgery},
  year = {2026},
  volume = {12},
  number = {11},
  pages = {175--175},
  doi = {10.21037/jss-2026-0240}
}
Kostenlos starten →