Operationally stable pH measurement in wastewater treatment using a photocatalytic self-cleaning electrode: Extended field evidence and uncertainty-oriented performance indicators

Abraham De Guzman, Elyson Keith Encarnacion

Acta IMEKO · 23.06.2026 · 15 (2) · 1-7

DOI: 10.21014/actaimeko.v15i2.2299 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta IMEKO, online erschienen am 23.06.2026. Verfasst von Abraham De Guzman und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study presents a 29-day field deployment of a HORIBA photocatalytic self-cleaning pH electrode for continuous wastewater monitoring in a full-scale treatment facility discharging into Laguna Lake, Philippines. The system generated 41,729 valid one-minute pH measurements between 11 December 2025 and 9 January 2026 with near-continuous uptime. The work focuses on uncertainty-oriented operational indicators derived from grouped observations, rather than on a complete formal Guide to the Expression of Uncertainty in Measurement (GUM) uncertainty budget. Statistical indicators included agreement with laboratory reference measurements, rolling median absolute deviation (MAD), grouped daily statistics, trend analysis, and distributional behavior. Laboratory reference measurements were obtained using a benchtop HORIBA pH meter, calibrated with buffer standards traceable to the National Institute of Standards and Technology (NIST). The results indicated low short-term variability, operationally stable variance behavior, preserved sensitivity to transient process events, and no evidence of uncontrolled sensor degradation during deployment. Type A and Type B uncertainty contributions were discussed qualitatively within a grouped-observation framework consistent with JCGM 100:2008 Appendix H.5. The observed temporal behavior was more consistent with gradual process evolution than with fouling-induced sensor instability. Future investigations include deployment in multiple scenarios for inter-electrode reproducibility, parallel experiments with controlled environments, and extended and multi-season temporal studies.

Zitieren

APA

Guzman, A. D., & Encarnacion, E. K. (2026). Operationally stable pH measurement in wastewater treatment using a photocatalytic self-cleaning electrode: Extended field evidence and uncertainty-oriented performance indicators. Acta IMEKO, 15(2), 1-7. https://doi.org/10.21014/actaimeko.v15i2.2299

Vancouver

Guzman AD, Encarnacion EK. Operationally stable pH measurement in wastewater treatment using a photocatalytic self-cleaning electrode: Extended field evidence and uncertainty-oriented performance indicators. Acta IMEKO. 2026;15(2):1-7. doi:10.21014/actaimeko.v15i2.2299

BibTeX
@article{Guzman2026,
  author = {Guzman, Abraham De and Encarnacion, Elyson Keith},
  title = {Operationally stable pH measurement in wastewater treatment using a photocatalytic self-cleaning electrode: Extended field evidence and uncertainty-oriented performance indicators},
  journal = {Acta IMEKO},
  year = {2026},
  volume = {15},
  number = {2},
  pages = {1--7},
  doi = {10.21014/actaimeko.v15i2.2299}
}
Kostenlos starten →