Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities
Physical Activity and Nutrition · 30.06.2026 · 30 (2) · 142-150
DOI: 10.20463/pan.2026.0032 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physical Activity and Nutrition, erschienen am 30.06.2026. Verfasst von Yoonyoung Jeong und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Sports Conditioning: Nutrition Biochemistry and Neuroscience, Department of Sport Science, College of Arts and Sports, University of Se…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 2.0). Gefördert von: Ministry of Education, National Research Foundation of Korea, Gimcheon University.
Abstract (Originalsprache)
[Purpose] We compared the heart rate (HR) measurement accuracy of commercially available PPG-based wearable devices against an ECG-based reference device across different exercise intensities.[Methods] Forty healthy young adults were recruited, and data from 37 participants were analyzed. HR was simultaneously measured using Polar H10 as the ECG-based reference and four PPG-based devices: Polar Verity Sense, Polar Pacer Pro, Apple Watch SE, and Galaxy Watch 6. Participants completed low-, moderate-, and high-intensity treadmill exercise. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), Lin’s concordance correlation coefficient (CCC), and Bland-Altman analysis.[Results] MAE and MAPE generally decreased as intensity increased. Significant differences in MAE were observed among the devices across intensities (p = 0.002). The Polar Verity Sense and Apple Watch SE showed significantly lower MAE than the Polar Pacer Pro and Galaxy Watch 6. During low-intensity exercise, the Apple Watch SE showed lower MAE than the Polar Pacer Pro and Galaxy Watch 6, whereas during high-intensity exercise, the Polar Verity Sense showed lower MAE than the Galaxy Watch 6. CCC values increased with exercise intensity and were consistently higher for the Polar Verity Sense and Apple Watch SE. Bland-Altman analysis showed narrower limits of agreement for the Apple Watch SE and Polar Verity Sense.[Conclusion] HR measurement accuracy differed substantially among the PPG-based wearable devices. The Apple Watch SE and Polar Verity Sense showed the highest agreement with the ECG-based reference, suggesting that wearable HR data should be interpreted based on device-specific validation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratory of Sports Conditioning: Nutrition Biochemistry and Neuroscience, Department of Sport Science, College of Arts and Sports, University of Seoul, Seoul, Republic of Korea.
- Institute of Health and Exercise for Urban Older Adults, University of Seoul, Seoul, Republic of Korea.
- R&D Center for Tailor‑Made QOL, University of Tsukuba, Tsukuba, Japan.
- International Institute for Integrative Sleep Medicine (WPI-IIIS), Tsukuba Institute for Advanced Research (TIAR), University of Tsukuba, Tsukuba, Japan.
- Department of Sports and Leisure Studies, Gimcheon University, Gimcheon, Republic of Korea. phdjwm@gmail.com.
- Laboratory of Sports Conditioning: Nutrition Biochemistry and Neuroscience, Department of Sport Science, College of Arts and Sports, University of Seoul, Seoul, Republic of Korea. haminseong@uos.ac.kr.
Förderung
- Ministry of Education
- National Research Foundation of Korea
- Gimcheon University
Zitieren
APA
Jeong, Y., Yoon, S., Lee, J., Jeong, W. M., & Ha, M. S. (2026). Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities. Physical Activity and Nutrition, 30(2), 142-150. https://doi.org/10.20463/pan.2026.0032
Vancouver
Jeong Y, Yoon S, Lee J, Jeong WM, Ha MS. Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities. Physical Activity and Nutrition. 2026;30(2):142-150. doi:10.20463/pan.2026.0032
BibTeX
@article{Jeong2026,
author = {Jeong, Yoonyoung and Yoon, Sehwa and Lee, Jaehee and Jeong, Woo-Min and Ha, Min-Seong},
title = {Comparison of heart rate measurement accuracy among commercially available photoplethysmography-based wearable devices across exercise intensities},
journal = {Physical Activity and Nutrition},
year = {2026},
volume = {30},
number = {2},
pages = {142--150},
doi = {10.20463/pan.2026.0032}
}
EUVIMED