Stability of pixel density in cone-beam computed tomographic images
Journal of Osaka Dental University · 01.10.2014 · 48 (2) · 97-102
DOI: 10.18905/jodu.48.2_97 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Osaka Dental University, erschienen am 01.10.2014. Verfasst von Yukako Nakashima und 2 weiteren Autor:innen.
Abstract (Originalsprache)
The use of cone-beam computed tomography (CBCT) is well accepted for dental diagnosis. However, there has not been adequate evaluation of the image quality that takes into consideration pixel density and dimensional integrity. As a result, some manufacturers claim without substantiation that their CBCT adequately displays CT numbers (CTN). Several authors have claimed that the same volume data can be obtained with CBCT as with multi-detector row computed tomography (MDCT). The density of CBCT images are not displayed as actual CTN because of structural limitations that include radiating condition, scatter radiation, dynamic range limitation of the detector, beam hardening, imaging parameters and object positioning. Another author has reported that the pixel density of CBCT can be used like CTN because there is linearity between pixel density in CBCT and the concentration of the materials. We also found a correlation between pixel density of CBCT and MDCT images. We took into consideration differences in the exposure condition and position of the object in the exposure field for the two image types. Our first goal was to check the stability of the density for the two image types. Then we clarified the correlation of the pixel density between the CBCT and MDCT images by exposing various materials and analyzing the results.
Schlagwörter
Zitieren
APA
Nakashima, Y., Yotsui, Y., & Shimizutani, K. (2014). Stability of pixel density in cone-beam computed tomographic images. Journal of Osaka Dental University, 48(2), 97-102. https://doi.org/10.18905/jodu.48.2_97
Vancouver
Nakashima Y, Yotsui Y, Shimizutani K. Stability of pixel density in cone-beam computed tomographic images. Journal of Osaka Dental University. 2014;48(2):97-102. doi:10.18905/jodu.48.2_97
BibTeX
@article{Nakashima2014,
author = {Nakashima, Yukako and Yotsui, Yoritaka and Shimizutani, Kimishige},
title = {Stability of pixel density in cone-beam computed tomographic images},
journal = {Journal of Osaka Dental University},
year = {2014},
volume = {48},
number = {2},
pages = {97--102},
doi = {10.18905/jodu.48.2_97}
}
EUVIMED