A novel method of furan determination in beer using gas chromatography with a flame ionization detector
KVASNY PRUMYSL · 2019
DOI: 10.18832/kp2019.65.174 ↗Kurz gefasst
Fachartikel in der Zeitschrift KVASNY PRUMYSL, 2019. Verfasst von Jana Olšovská und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Furan is considered to have a potential carcinogenic effect on human health. Its presence was found in a number of foods including beer. Thus, it is necessary to have a fast and routine method for its determination. This study describes a procedure meeting these requirements using the head-space technique associated with gas chromatography. A capillary column with the length of 60 m, inner diameter of 0.32 mm and the film thickness of 0.25 μm was used for the determination of furan. A flame-ionization detector (FID) was used for the detection. The characteristics of the procedure are presented.
Zitieren
APA
Olšovská, J., Horák, T., & Štěrba, K. (2019). A novel method of furan determination in beer using gas chromatography with a flame ionization detector. KVASNY PRUMYSL. https://doi.org/10.18832/kp2019.65.174
Vancouver
Olšovská J, Horák T, Štěrba K. A novel method of furan determination in beer using gas chromatography with a flame ionization detector. KVASNY PRUMYSL. 2019. doi:10.18832/kp2019.65.174
BibTeX
@article{Olovsk2019,
author = {Olšovská, Jana and Horák, Tomáš and Štěrba, Karel},
title = {A novel method of furan determination in beer using gas chromatography with a flame ionization detector},
journal = {KVASNY PRUMYSL},
year = {2019},
doi = {10.18832/kp2019.65.174}
}