EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

The Fundus Red Reflex Test Using a Smartphone Showed High Accuracy Compared to an Autorefractor to Detect Refractive Errors in Children

Muhammad Syauqie, Lutfah Rif'ati, Wienta Diarsvitri, Sutanto Priyo Hastono, Kemal Nazaruddin Siregar, Nila Djuwita Farieda Moeloek

Journal of Ophthalmic and Vision Research · 06.08.2026 · 21 · 1-9

DOI: 10.18502/jovr.v21.18164 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Ophthalmic and Vision Research, online erschienen am 06.08.2026. Verfasst von Muhammad Syauqie und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Faculty of Medicine, Andalas University, Padang, Indonesia. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To determine the accuracy of the fundus red reflex test using a smartphone and autorefractor to detect refractive errors in children.

Methods

This cross-sectional study included 1302 eyes of 651 school-age children aged 6- 13 years with visual acuity worse than 6/9 in at least one eye, examined using the Snellen chart. Study participants were selected using purposive sampling, after which non-cycloplegic refractive measurement was performed using two instruments: a smartphone-based application (MD EyeCare) and an autorefractor. The detection of refractive errors using the online platform was compared with autorefraction as the reference standard.

Results

Our study found that 46.10% of participants with abnormal red reflex had significant myopia, 41.20% had nonsignificant refractive errors, and 12.70% had significant hypermetropia. The mean difference between the smartphone-based application and the autorefractor was ≤−3.00 D (range: −2.88 to −2.63), with a sensitivity of 0.89 to 0.90 and a specificity of 0.86 to 0.87 for significant myopia, and the area under the curve was between 0.94 and 0.97. The mean difference between the smartphone-based application and the autorefractor was ≥+1.00 D (range: +0.13 to +0.38), with a sensitivity of 0.76 and a specificity of 0.97 for significant hypermetropia, and the area under the curve was between 0.88 and 0.96.

Conclusion

A smartphone-based application could be used as a screening tool for refractive errors in children, especially in resource-limited settings.

Schlagwörter

Stichwörter der Autor:innen

ChildrenRefractive ErrorsSensitivitySmartphone-based ApplicationSpecificityAutorefractor

Autor:innen und Institutionen

  • Department of Ophthalmology, Faculty of Medicine, Andalas University, Padang, Indonesia.
  • Center for Preclinical and Clinical Medical Research, National Research and Innovation Agency, Jakarta, Indonesia.
  • Department of Community Medicine, Faculty of Medicine, Hang Tuah University, Surabaya, Indonesia.
  • Department of Biostatistics and Population Studies, Faculty of Public Health, Universitas Indonesia, Depok, Indonesia.
  • Department of Ophthalmology, Faculty of Medicine, University of Indonesia / Cipto Mangunkusumo Hospital, Jakarta, Indonesia.

Zitieren

APA

Syauqie, M., Rif'ati, L., Diarsvitri, W., Hastono, S. P., Siregar, K. N., & Moeloek, N. D. F. (2026). The Fundus Red Reflex Test Using a Smartphone Showed High Accuracy Compared to an Autorefractor to Detect Refractive Errors in Children. Journal of Ophthalmic and Vision Research, 21, 1-9. https://doi.org/10.18502/jovr.v21.18164

Vancouver

Syauqie M, Rif'ati L, Diarsvitri W, Hastono SP, Siregar KN, Moeloek NDF. The Fundus Red Reflex Test Using a Smartphone Showed High Accuracy Compared to an Autorefractor to Detect Refractive Errors in Children. Journal of Ophthalmic and Vision Research. 2026;21:1-9. doi:10.18502/jovr.v21.18164

BibTeX
@article{Syauqie2026,
  author = {Syauqie, Muhammad and Rif'ati, Lutfah and Diarsvitri, Wienta and Hastono, Sutanto Priyo and Siregar, Kemal Nazaruddin and Moeloek, Nila Djuwita Farieda},
  title = {The Fundus Red Reflex Test Using a Smartphone Showed High Accuracy Compared to an Autorefractor to Detect Refractive Errors in Children},
  journal = {Journal of Ophthalmic and Vision Research},
  year = {2026},
  volume = {21},
  pages = {1--9},
  doi = {10.18502/jovr.v21.18164}
}
Kostenlos starten →