Simple “Digital Phantom” for Testing Attenuation-Imaging Methods in Optical Coherence Tomography
Journal of Biomedical Photonics & Engineering · 26.04.2024 · 020302
DOI: 10.18287/jbpe24.10.020302 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomedical Photonics & Engineering, online erschienen am 26.04.2024. Verfasst von Alexander A. Sovetsky und 4 weiteren Autor:innen. Die Arbeit wurde bisher 6-mal zitiert (Crossref).
Abstract (Originalsprache)
The manuscript describes a simple but realistic and efficient method for generation of OCT scans with fairly correct accounting for optical signal attenuation dominated by scattering. The strong domination of scattering in the optical attenuation coefficient (OAC) is typical of OCT where the illuminating-beam wavelength is intentionally chosen in optical transparently windows with minimal absorption. At the same time the scattering-dominated OAC may strongly differ for various types/states of biological tissues, so that the diagnostic value of OCT examinations can be significantly increased due to spatially-resolved estimation of OAC. However, the results of OAC-reconstruction may be strongly degraded by the “speckle noise” intrinsic to OCT scans, various measurement noises, and some other factors. Verification of OAC-reconstruction accuracy in physical experiments is challenging since controllable variation in the parameters of physical phantoms and OCT systems is often difficult or even impossible. In view of this, realistic numerical simulations may open unprecedented possibilities for testing and comparison of various OAC methods in highly controllable and flexibly variable conditions. Here, we describe a method enabling generation of realistic “digital phantoms” and present instructive examples demonstrating their usefulness for testing OAC-reconstruction approaches.
Zitieren
APA
Sovetsky, A. A., Matveyev, A. L., Matveev, L. A., Chizhov, P. A., & Zaitsev, V. Y. (2024). Simple “Digital Phantom” for Testing Attenuation-Imaging Methods in Optical Coherence Tomography. Journal of Biomedical Photonics & Engineering, 020302. https://doi.org/10.18287/jbpe24.10.020302
Vancouver
Sovetsky AA, Matveyev AL, Matveev LA, Chizhov PA, Zaitsev VY. Simple “Digital Phantom” for Testing Attenuation-Imaging Methods in Optical Coherence Tomography. Journal of Biomedical Photonics & Engineering. 2024:020302. doi:10.18287/jbpe24.10.020302
BibTeX
@article{Sovetsky2024,
author = {Sovetsky, Alexander A. and Matveyev, Alexander L. and Matveev, Lev A. and Chizhov, Peter A. and Zaitsev, Vladimir Y.},
title = {Simple “Digital Phantom” for Testing Attenuation-Imaging Methods in Optical Coherence Tomography},
journal = {Journal of Biomedical Photonics & Engineering},
year = {2024},
pages = {020302},
doi = {10.18287/jbpe24.10.020302}
}