Revision Total Knee Arthroplasty Using an Implant-Agnostic, Joint-Line Based, Gap-Balanced Technique
Medical Research Archives · 31.05.2026 · 14 (5)
DOI: 10.18103/mra.v14i5.7550 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Research Archives, online erschienen am 31.05.2026. Verfasst von Spencer J Montgomery und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
Instability is a significant cause of failure following revision total knee arthroplasty and results in the lowest patient satisfaction. This study validates, under controlled cadaveric parameters, a joint-line based, gap-balancing technique for performing revision total knee arthroplasty using an implant-agnostic workflow. Materials and Methods: A cadaveric feasibility study was completed with eight lower extremities previously prepared for total knee arthroplasty in kinematic alignment undergoing revision total knee arthroplasty to mechanical alignment with a joint-line based, gap-balanced workflow. A step-by-step guide for this technique was documented and performed, and accuracy of the approach was quantified in all cases.
Results
After bone preparation, the measured gaps were all within 1mm of the desired measurement (medial extension 0.38mm, lateral extension 0.63mm, medial flexion 0.75mm, and lateral flexion 0.25mm), accepting the pre-calculated implant and augment sizes. There were no adjustments needed to bone cuts or soft tissues in any specimen. Of interest, 10mm posterolateral augments were required in all specimens.
Conclusions
The technique described in this study showed promising accuracy in a series of cadaveric specimens. With additional clinical testing, this approach to revision total knee arthroplasty could prove to reduce the need for re-revision and improve patient satisfaction.
Zitieren
APA
Montgomery, S. J., Sikes, J. H., & Aparicio, H. A. (2026). Revision Total Knee Arthroplasty Using an Implant-Agnostic, Joint-Line Based, Gap-Balanced Technique. Medical Research Archives, 14(5). https://doi.org/10.18103/mra.v14i5.7550
Vancouver
Montgomery SJ, Sikes JH, Aparicio HA. Revision Total Knee Arthroplasty Using an Implant-Agnostic, Joint-Line Based, Gap-Balanced Technique. Medical Research Archives. 2026;14(5). doi:10.18103/mra.v14i5.7550
BibTeX
@article{Montgomery2026,
author = {Montgomery, Spencer J and Sikes, James H and Aparicio, Humberto A.},
title = {Revision Total Knee Arthroplasty Using an Implant-Agnostic, Joint-Line Based, Gap-Balanced Technique},
journal = {Medical Research Archives},
year = {2026},
volume = {14},
number = {5},
doi = {10.18103/mra.v14i5.7550}
}
EUVIMED