EUVIMEDEUVIMED entdecken
Studienprotokoll

Validation of a Low-Dose and Time-Efficient 13N-Ammonia PET/CT Protocol

Yoko Kaimoto, Kenji Fukushima, Atsushi Yamamoto, Risa Imakado, Risako Nakao, Yui Maekawa, Akiko Sakai, Mitsuru Momose, Michinobu Nagao, Junichi Yamaguchi, Shuji Sakai

Annals of Nuclear Cardiology · 2026 · 12 (1) · 12-17

DOI: 10.17996/anc.25-00006 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Annals of Nuclear Cardiology, 2026. Verfasst von Yoko Kaimoto und 10 weiteren Autor:innen. Die Arbeit nennt 15 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Although 13 N-Ammonia (NH3) positron emission tomography (PET) has been increasingly applied in clinical practice, its long examination time remains a limitation, primarily due to the required interval for sufficient radioactive decay.This study aimed to validate a low-dose rest scan to reduce the required interval and shorten the overall examination time.

Methods

This prospective study enrolled 24 patients (age 65±12, 15 male) with known or suspected coronary artery disease.To validate the low-dose approach, each patient underwent two separate rest scans consecutively: a low-dose scan (approximately one-third the standard dose) and a standard-dose scan (555 to 740 MBq), immediately followed by standard pharmacological stress scan.PET parameters, including end-diastolic/systolic volumes (EDV/ESV, mL), ejection fraction (EF, %), summed rest score (SRS), and rest myocardial blood flow (MBF, mL/g/min), were compared between the two rest protocols.

Results

EF, EDV, ESV, SRS, and MBF showed significant correlation between low and standard-dose (r=0.99,0. 99, 0.99, 0.99, and 0.93 for EF, EDV, ESV, SRS, and MBF; p

Zitieren

APA

Kaimoto, Y., Fukushima, K., Yamamoto, A., Imakado, R., Nakao, R., Maekawa, Y., Sakai, A., Momose, M., Nagao, M., Yamaguchi, J., & Sakai, S. (2026). Validation of a Low-Dose and Time-Efficient 13N-Ammonia PET/CT Protocol. Annals of Nuclear Cardiology, 12(1), 12-17. https://doi.org/10.17996/anc.25-00006

Vancouver

Kaimoto Y, Fukushima K, Yamamoto A, Imakado R, Nakao R, Maekawa Y, et al. Validation of a Low-Dose and Time-Efficient 13N-Ammonia PET/CT Protocol. Annals of Nuclear Cardiology. 2026;12(1):12-17. doi:10.17996/anc.25-00006

BibTeX
@article{Kaimoto2026,
  author = {Kaimoto, Yoko and Fukushima, Kenji and Yamamoto, Atsushi and Imakado, Risa and Nakao, Risako and Maekawa, Yui and Sakai, Akiko and Momose, Mitsuru and Nagao, Michinobu and Yamaguchi, Junichi and Sakai, Shuji},
  title = {Validation of a Low-Dose and Time-Efficient 13N-Ammonia PET/CT Protocol},
  journal = {Annals of Nuclear Cardiology},
  year = {2026},
  volume = {12},
  number = {1},
  pages = {12--17},
  doi = {10.17996/anc.25-00006}
}
Kostenlos starten →