Comparison of Taiwanese and European Calibration Factors for Heart-to-Mediastinum Ratio in Multicenter 123I-mIBG Phantom Studies

Koichi Okuda, Kenichi Nakajima, Guang-Uei Hung, Hao-Ting Wu, Derk O Verschure, Hein J Verberne, Chiemi Kitamura

Annals of Nuclear Cardiology · 2023 · 9 (1) · 54-60

DOI: 10.17996/anc.23-00006 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Nuclear Cardiology, 2023. Verfasst von Koichi Okuda und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Science, Hirosaki University Graduate School of Health Sciences, Aomori, Japan. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: JSPS KAKENHI, the National Science and Technology Council of Taiwan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Cross-calibration of 123I-labeled meta-iodobenzylguanidine (mIBG) myocardial-derived indices is essential to extrapolate findings from several clinical centers. Here, we conducted a phantom study to generate conversion coefficients for the calibration of heart-to-mediastinum ratios and compare them between Taiwan and Europe.

Methods

We used an acrylic phantom dedicated to 123I-mIBG planar imaging to calculate the conversion coefficients of 136 phantom images derived from 36 Taiwanese institutions. A European phantom image database including 191 images from 27 institutions was used. Conversion coefficients were categorized into five collimator types: low-energy (LE) high-resolution (LEHR), LE general-purpose (LEGP), extended LEGP (ELEGP), medium-energy (ME) GP (MEGP), and ME low-penetration (MELP) collimators.

Results

The conversion coefficients were 0.53 ± 0.039, 0.59 ± 0.032, 0.79 ± 0.032, 0.96 ± 0.038, and 0.99 ± 0.050 for LEHR, LEGP, ELEGP, MEGP, and MELP collimators, respectively. The Taiwanese and European conversion coefficients for the LEHR, LEGP, and MELP collimators did not significantly differ. The coefficient of variation was slightly higher for the Taiwanese than the European conversion coefficients (3.7%-7.5% vs. 2.3%-5.6%).

Conclusions

We calculated conversion coefficients for various types of collimators used in Taiwan using a 123I-mIBG phantom. In general, the Taiwanese and European conversion coefficients were comparable. These findings further corroborated and highlighted the need for 123I-mIBG standardization using the phantom-determined conversion coefficients.

Schlagwörter

Stichwörter der Autor:innen

123I-mIBGCalibrationCollimatorHeart-to-mediastinum ratioPhantom

Autor:innen und Institutionen

  • Department of Radiation Science, Hirosaki University Graduate School of Health Sciences, Aomori, Japan.
  • Department of Physics, Kanazawa Medical University, Ishikawa, Japan.
  • Department of Functional Imaging and Artificial Intelligence, Graduate School of Advanced Preventive Medical Sciences, Kanazawa University, Ishikawa, Japan.
  • Department of Nuclear Medicine, Chang Bing Show Chwan Memorial Hospital, Changhua, Taiwan.
  • Department of Radiology and Nuclear Medicine, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Department of Cardiology, Zaans Medical Center, Zaandam, The Netherlands.

Förderung

  • JSPS KAKENHI
  • the National Science and Technology Council of Taiwan

Zitieren

APA

Okuda, K., Nakajima, K., Hung, G. U., Wu, H. T., Verschure, D. O., Verberne, H. J., & Kitamura, C. (2023). Comparison of Taiwanese and European Calibration Factors for Heart-to-Mediastinum Ratio in Multicenter 123I-mIBG Phantom Studies. Annals of Nuclear Cardiology, 9(1), 54-60. https://doi.org/10.17996/anc.23-00006

Vancouver

Okuda K, Nakajima K, Hung GU, Wu HT, Verschure DO, Verberne HJ, et al. Comparison of Taiwanese and European Calibration Factors for Heart-to-Mediastinum Ratio in Multicenter 123I-mIBG Phantom Studies. Annals of Nuclear Cardiology. 2023;9(1):54-60. doi:10.17996/anc.23-00006

BibTeX
@article{Okuda2023,
  author = {Okuda, Koichi and Nakajima, Kenichi and Hung, Guang-Uei and Wu, Hao-Ting and Verschure, Derk O and Verberne, Hein J and Kitamura, Chiemi},
  title = {Comparison of Taiwanese and European Calibration Factors for Heart-to-Mediastinum Ratio in Multicenter 123I-mIBG Phantom Studies},
  journal = {Annals of Nuclear Cardiology},
  year = {2023},
  volume = {9},
  number = {1},
  pages = {54--60},
  doi = {10.17996/anc.23-00006}
}
Kostenlos starten →