EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Evaluation of the physicomechanical properties of magnesium alloy herbert screws

Maxim N. Semenisty, Yuliya S. Lukina, Leonid L. Bionyshev-Abramov, Alexander A. Ochkurenko, Anatoly V. Ovsyankin, Nina V. Zykova

N.N. Priorov Journal of Traumatology and Orthopedics · 04.04.2026 · 33 (2) · 342-351

DOI: 10.17816/vto699990 ↗

Kurz gefasst

Fachartikel in der Zeitschrift N.N. Priorov Journal of Traumatology and Orthopedics, online erschienen am 04.04.2026. Verfasst von Maxim N. Semenisty und 5 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Magnesium-based biodegradable alloys used for fixation devices exhibit substantially lower mechanical strength compared with titanium counterparts, due to their strength characteristics being closer to those of bone tissue. Although magnesium screws have demonstrated favorable clinical outcomes, their mechanical properties have not been sufficiently evaluated. AIM: The work aimed to perform a comparative analysis of the physicomechanical properties of Herbert screws made of bioresorbable MgYREZr magnesium alloy produced by Russian and international manufacturers. METHODS: This was a prospective, pilot, multicenter, experimental study of Herbert screws made of MgYREZr alloy manufactured by Syntellix AG (Hannover, Germany) and Osteo-Sibear LLC (Moscow, Russia), with a consecutive sample of eleven screws. The study was experimental and involved the accumulation of both quantitative and qualitative data. The control included both descriptive assessment and precise measurements with before/after comparison. During testing, processes simulating displacement of bone structures were modeled. Herbert screws were fixed in polyurethane blocks manufactured according to the ASTM F1839 standard and tested using a universal testing machine (Walter+Bai AG LFM-50). The setup was used to simulate pull-out, shear failure, and stripping of the screw head. RESULTS: The mean pull-out force for the polymer block fixed with the Osteo-Sibear screw was 301.33 N, whereas 160 N was required to disrupt fixation with the Syntellix AG screw. To induce displacement of the polymer block fixed with the Osteo-Sibear screw, a force of approximately 298.5 N was required, whereas the corresponding fixation with the Syntellix AG screw failed at 295 N. The mean torque required to cause screw head slot stripping was 0.23 Nm for the Osteo-Sibear screw and 0.36 Nm for the Syntellix AG screw. CONCLUSION: The Russian Herbert screw demonstrated superior fixation strength in pull-out and shear tests. Under torsional loading, the screw head slot failed, with the screw body remaining intact.

Zitieren

APA

Semenisty, M. N., Lukina, Y. S., Bionyshev-Abramov, L. L., Ochkurenko, A. A., Ovsyankin, A. V., & Zykova, N. V. (2026). Evaluation of the physicomechanical properties of magnesium alloy herbert screws. N.N. Priorov Journal of Traumatology and Orthopedics, 33(2), 342-351. https://doi.org/10.17816/vto699990

Vancouver

Semenisty MN, Lukina YS, Bionyshev-Abramov LL, Ochkurenko AA, Ovsyankin AV, Zykova NV. Evaluation of the physicomechanical properties of magnesium alloy herbert screws. N.N. Priorov Journal of Traumatology and Orthopedics. 2026;33(2):342-351. doi:10.17816/vto699990

BibTeX
@article{Semenisty2026,
  author = {Semenisty, Maxim N. and Lukina, Yuliya S. and Bionyshev-Abramov, Leonid L. and Ochkurenko, Alexander A. and Ovsyankin, Anatoly V. and Zykova, Nina V.},
  title = {Evaluation of the physicomechanical properties of magnesium alloy herbert screws},
  journal = {N.N. Priorov Journal of Traumatology and Orthopedics},
  year = {2026},
  volume = {33},
  number = {2},
  pages = {342--351},
  doi = {10.17816/vto699990}
}
Kostenlos starten →