EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Validation of a Russian automated CT perfusion processing system for ischemic stroke

Ilya L. Gubskiy, Mikhail M. Beregov, Nikolai E. Staroverov, Ivan A. Larionov, Kirill Yu. Kazachkov, Andrey P. Stepanchenko, Valentin A. Nechaev, Nataliya A. Marskaya, Nikolay A. Shamalov

Journal of Clinical Practice · 09.04.2026 · 17 (1) · 50-61

DOI: 10.17816/clinpract703530 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Practice, online erschienen am 09.04.2026. Verfasst von Ilya L. Gubskiy und 8 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Automated computed tomography (CT) perfusion analysis systems are critical for patient selection for reperfusion therapy in acute ischemic stroke. Systems such as RAPID (iSchemaView/RapidAI, USA) and Olea Sphere (Olea Medical, France) are widely used in clinical practice; however, a validated domestic system for automated CT perfusion processing is lacking in the Russian Federation. Among similar systems that have a registration certificate in the Russian Federation, Vitrea (Canon Medical Systems, USA, Japan) is available in clinical practice. AIM: To validate the Russian system ArchiMed PRO Chronos by comparing it with the Vitrea (Canon Medical Systems, USA, Japan) software for quantitative assessment of ischemic core and critical hypoperfusion area. METHODS: A retrospective study included 52 patients with acute ischemic stroke caused by large vessel occlusion and symptom onset within 24 hours. Correlation between systems was assessed for ischemic core volume and critical hypoperfusion area determination. RESULTS: Strong correlation was found between systems for ischemic core volume (r=0.95) and critical hypoperfusion area (r=0.85). After excluding studies with inadequate segmentation, correlation improved to r=0.98 and r=0.93, respectively. Adequate segmentation was achieved in 84.6% of cases for both systems. No statistically significant differences between systems were detected (p 0.05). CONCLUSION: The ArchiMed PRO Chronos system demonstrated excellent agreement with Vitrea and can be used for quantitative CT perfusion analysis and patient selection for reperfusion therapy.

Zitieren

APA

Gubskiy, I. L., Beregov, M. M., Staroverov, N. E., Larionov, I. A., Kazachkov, K. Y., Stepanchenko, A. P., Nechaev, V. A., Marskaya, N. A., & Shamalov, N. A. (2026). Validation of a Russian automated CT perfusion processing system for ischemic stroke. Journal of Clinical Practice, 17(1), 50-61. https://doi.org/10.17816/clinpract703530

Vancouver

Gubskiy IL, Beregov MM, Staroverov NE, Larionov IA, Kazachkov KY, Stepanchenko AP, et al. Validation of a Russian automated CT perfusion processing system for ischemic stroke. Journal of Clinical Practice. 2026;17(1):50-61. doi:10.17816/clinpract703530

BibTeX
@article{Gubskiy2026,
  author = {Gubskiy, Ilya L. and Beregov, Mikhail M. and Staroverov, Nikolai E. and Larionov, Ivan A. and Kazachkov, Kirill Yu. and Stepanchenko, Andrey P. and Nechaev, Valentin A. and Marskaya, Nataliya A. and Shamalov, Nikolay A.},
  title = {Validation of a Russian automated CT perfusion processing system for ischemic stroke},
  journal = {Journal of Clinical Practice},
  year = {2026},
  volume = {17},
  number = {1},
  pages = {50--61},
  doi = {10.17816/clinpract703530}
}
Kostenlos starten →