A New Criterion for Shear Wave Elastometric Assessment Using Modulus of Stiffness Difference between Object and Environment
Sovremennye tehnologii v medicine · 31.10.2022 · 14 (5) · 5
DOI: 10.17691/stm2022.14.5.01 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sovremennye tehnologii v medicine, online erschienen am 31.10.2022. Verfasst von I Yu Demin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Associate Professor, Department of Acoustics, Faculty of Radiophysics; National Research Lobachevsky State University of Nizhni Novgorod, 23 Prospekt…. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques. Die Arbeit wurde bisher 3-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Ultrasound shear wave elastography is a modern method that is based on measuring the shear wave velocity making it possible to determine stiffness of soft biological tissues at an arbitrary point (point elastography) or to construct a two-dimensional color image with a subsequent point measurement of stiffness (two-dimensional elastography) and therefore to compare the stiffness of an object with a medium or objects with each other. The aim of the study is to develop a new criterion for the comparative assessment of objects with different stiffness during shear wave elastometry: modulus of stiffness difference between object and environment. Materials and Methods: Using the original technology of building two-dimensional color elastogram, point and two-dimensional shear wave elastography were performed using linear sensors on commercial ultrasound scanners: Aixplorer (SuperSonic Imagine, France), Acuson S2000 (Siemens, Germany), and Verasonics acoustic system (Verasonics Inc., USA) with an open architecture to determine the stiffness values of focal inclusions and compare them with each other with the help of a new comparative elastomeric assessment criterion: modulus of stiffness difference between object and environment. First, the accuracy of the scanners under test was compared on a calibrated Elasticity QA Phantom, model 049 (Computerized Imaging Reference Systems Company, USA) with a known stiffness of various inclusions and thereafter on an uncalibrated BP1901 phantom (Blue Phantom, USA) with unknown stiffness of inclusions. The obtained values were compared to determine the influence of subjective factors on the measurement results.
Results
To assess the stiffness of the foci and compare the values with each other taking into account the rigidity of the environment, it is proposed to use a new criterion for the comparative assessment - the modulus of stiffness difference between focus and environment, which quantitatively characterizes the difference between these values. According to this criterion, all three ultrasound scanners have been established to show high and comparable accuracy in determining the stiffness of inclusions within the homogeneous medium in the experiments on phantoms. Two-dimensional shear wave elastography has revealed the effect of the control volume size and the correctness of the color scale setting, especially in the heterogeneous objects, on the results of elastometry. Methodological techniques to reduce the influence of subjective factors have also been proposed.
Conclusion
The study has showed the possibility of using the modulus of stiffness difference between object and environment as a new criterion for comparative assessment of objects in shear wave elastometry taking into account stiffness of the environment. To reduce operator-dependence, it is necessary to take into consideration both the way of realizing elastometry (point or two-dimensional color elastography) and a number of other methodological factors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Associate Professor, Department of Acoustics, Faculty of Radiophysics; National Research Lobachevsky State University of Nizhni Novgorod, 23 Prospekt Gagarina, Nizhny Novgorod, 603950, Russia.
- Head of the Department of Radiation Diagnostics; Volga District Medical Center of the Federal Medical and Biological Agency of Russia, 2 Nizhne-Volzhskaya naberezhnaya, Nizhny Novgorod, 603001, Russia.
- PhD Student, Faculty of Radiophysics; National Research Lobachevsky State University of Nizhni Novgorod, 23 Prospekt Gagarina, Nizhny Novgorod, 603950, Russia.
- Professor, Head of the Department of Radiation Diagnostics, Faculty of Advanced Medical Training; Privolzhsky Research Medical University, 10/1 Minin and Pozharsky Square, Nizhny Novgorod, 603005, Russia.
Zitieren
APA
Demin, I. Y., Rykhtik, P. I., Spivak, А. E., & Safonov, D. V. (2022). A New Criterion for Shear Wave Elastometric Assessment Using Modulus of Stiffness Difference between Object and Environment. Sovremennye tehnologii v medicine, 14(5), 5. https://doi.org/10.17691/stm2022.14.5.01
Vancouver
Demin IY, Rykhtik PI, Spivak АE, Safonov DV. A New Criterion for Shear Wave Elastometric Assessment Using Modulus of Stiffness Difference between Object and Environment. Sovremennye tehnologii v medicine. 2022;14(5):5. doi:10.17691/stm2022.14.5.01
BibTeX
@article{Demin2022,
author = {Demin, I Yu and Rykhtik, P I and Spivak, А E and Safonov, D V},
title = {A New Criterion for Shear Wave Elastometric Assessment Using Modulus of Stiffness Difference between Object and Environment},
journal = {Sovremennye tehnologii v medicine},
year = {2022},
volume = {14},
number = {5},
pages = {5},
doi = {10.17691/stm2022.14.5.01}
}