Designing and evaluating analytical parameters to adapt siemens urinary creatinine enzymatic method to open system analysers

Mina Ashraf, McNeice Leah, Banukumar Shanmugam, Vazquez Santiago

Open Journal of Analytical and Bioanalytical Chemistry · 2020

DOI: 10.17352/ojabc.000021 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Open Journal of Analytical and Bioanalytical Chemistry, 2020. Verfasst von Mina Ashraf und 3 weiteren Autor:innen. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Urinary creatinine is measured to assess kidney function and also as part of sample validity testing in drugs of abuse.Creatinine methods based on alkaline picrate Jaffé's reaction require extra cleaning on chemistry analysers to minimise any interference from picric acid and sodium hydroxide on other reagents on board.Enzymatic methods reagents are not as invasive and more specifi c.Siemens enzymatic method is more sensitive and specifi c when compared to Thermo Fisher alkaline picrate method but there were no available analytical parameters to setup the Siemens enzymatic method on Beckman-Coulter AU5800 analyser as an open system analyser.Emulating setup parameters from Siemens chemistry analysers did not work.The analytical parameters were developed through systematic testing using different settings to adopt and optimise the method.A full correlation was performed using the developed parameters with alkaline picrate method.The Deming regression weighted analysis for 438 samples showed a good correlation at a 95% confi dence interval.Slope is 1.029 to 1.041, Y-intercept when X=0.0 is -0.9156 to -0.7481 and correlation coeffi cient (r) is 0.998.Alkaline picrate method Mean and SD is 12.059 and 7.248 respectively and for the Enzymatic method is 11.653 and 7.504 respectively.Many interferences from drugs and other substances are eliminated when using the enzymatic method and the stability of other reagents on-board improved because the reagents used in the enzymatic method are less invasive.

Zitieren

APA

Ashraf, M., Leah, M., Shanmugam, B., & Santiago, V. (2020). Designing and evaluating analytical parameters to adapt siemens urinary creatinine enzymatic method to open system analysers. Open Journal of Analytical and Bioanalytical Chemistry. https://doi.org/10.17352/ojabc.000021

Vancouver

Ashraf M, Leah M, Shanmugam B, Santiago V. Designing and evaluating analytical parameters to adapt siemens urinary creatinine enzymatic method to open system analysers. Open Journal of Analytical and Bioanalytical Chemistry. 2020. doi:10.17352/ojabc.000021

BibTeX
@article{Ashraf2020,
  author = {Ashraf, Mina and Leah, McNeice and Shanmugam, Banukumar and Santiago, Vazquez},
  title = {Designing and evaluating analytical parameters to adapt siemens urinary creatinine enzymatic method to open system analysers},
  journal = {Open Journal of Analytical and Bioanalytical Chemistry},
  year = {2020},
  doi = {10.17352/ojabc.000021}
}
Kostenlos starten →