Study on the Impact of Different Tube Voltages Combined with Organ Dose Modulation on Cerebral Perfusion Image Quality and Ocular Lens Radiation Dose

Wentao MA, Yongxian ZHANG, Senlin GUO, Yunfu LIU, Tianliang KANG

CT Lilun yu yingyong yanjiu · 2026-01 · 35 (1) · 102-111

DOI: 10.15953/j.ctta.2024.337 ↗

Kurz gefasst

Fachartikel in der Zeitschrift CT Lilun yu yingyong yanjiu, 2026-01. Verfasst von Wentao MA und 4 weiteren Autor:innen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To investigate the effect of using organ dose modulation (ODM) technology at different tube voltages on image quality and eye lens radiation dose in brain CT perfusion (CTP).

Methods

Based on a clinical CTP scanning protocol, five tube voltages (70, 80, 100, 120, and 140 kV) were used to scan a Catphan phantom and a fresh, isolated human head specimen with three tube current modulation modes: without dose modulation technology (Manual mode), with smart tube current modulation technology (Smart mA mode), and with organ dose modulation technology (ODM mode). A long rod ionization chamber was fixed at a consistent position in front of the right eye lens. Each parameter combination was scanned nine times, and the average eye lens dose (Dav) was recorded. The modulation transfer function (MTF) of the CTP528 high-contrast resolution module in the Catphan phantom was measured. In the images of the head specimen, the heads of the caudate nucleus, lenticular nucleus, and thalamus were selected as signal regions at the basal ganglia level, and the lateral ventricle, anterior limb of the internal capsule, and posterior limb of the internal capsule were selected as respective background regions to measure the contrast-to-noise ratio (CNR). The same level of brain tissue was segmented to measure three texture feature parameters: contrast, correlation, and difference variance. Two-factor analysis of variance was used to compare MTF and CNR, and multi-factor non-parametric analysis of variance was used to compare eye lens dose and texture parameters.

Results

The differences in MTF across the different tube voltages were statistically significant, with MTF values positively correlated with tube voltage above 80 kV. However, there was no significant difference between different tube-current modulation modes. The CNR of each signal and background region showed significant differences among different tube voltages. The CNR for the caudate nucleus head to lateral ventricle was the largest, at 120 kV. The CNR for the lenticular nucleus/thalamus were negatively correlated with tube voltage. There were no significant differences in CNR among different tube-current modulation modes. No texture parameter was significantly associated with the tube current modulation mode. There were significant differences in Dav values with respect to both different tube voltages and tube current modulation modes. When the tube voltage was below 120 kV, the eye lens dose was negatively correlated with tube voltage, and the ODM technology reduced the eye lens dose by approximately 20% at each tube voltage.

Conclusion

Employing 80~100 kV CTP scanning with organ dose modulation technology can balance image parameters and reduce radiation dose to the eye lens while ensuring adequate image quality, as recommended for clinical use.

Schlagwörter

brain perfusion ctorgan dose modulationlensimage quality

Zitieren

APA

MA, W., ZHANG, Y., GUO, S., LIU, Y., & KANG, T. (2026). Study on the Impact of Different Tube Voltages Combined with Organ Dose Modulation on Cerebral Perfusion Image Quality and Ocular Lens Radiation Dose. CT Lilun yu yingyong yanjiu, 35(1), 102-111. https://doi.org/10.15953/j.ctta.2024.337

Vancouver

MA W, ZHANG Y, GUO S, LIU Y, KANG T. Study on the Impact of Different Tube Voltages Combined with Organ Dose Modulation on Cerebral Perfusion Image Quality and Ocular Lens Radiation Dose. CT Lilun yu yingyong yanjiu. 2026;35(1):102-111. doi:10.15953/j.ctta.2024.337

BibTeX
@article{MA2026,
  author = {MA, Wentao and ZHANG, Yongxian and GUO, Senlin and LIU, Yunfu and KANG, Tianliang},
  title = {Study on the Impact of Different Tube Voltages Combined with Organ Dose Modulation on Cerebral Perfusion Image Quality and Ocular Lens Radiation Dose},
  journal = {CT Lilun yu yingyong yanjiu},
  year = {2026},
  volume = {35},
  number = {1},
  pages = {102--111},
  doi = {10.15953/j.ctta.2024.337}
}
Kostenlos starten →