Deep Learning Image Reconstruction to Improve Computed Tomography Image Quality of the Phantom with Standard Liver Density

Zhijie PAN, Ling LIU, Qingyao LI, Tingting QU, Shuai ZHANG, Xueqian XIE

CT Lilun yu yingyong yanjiu · 2025-07 · 34 (4) · 677-685

DOI: 10.15953/j.ctta.2024.056 ↗

Kurz gefasst

Fachartikel in der Zeitschrift CT Lilun yu yingyong yanjiu, 2025-07. Verfasst von Zhijie PAN und 5 weiteren Autor:innen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to compare the quality of reconstructed images by deep learning image reconstruction (DLIR) and adaptive statistical iterative reconstruction-V (ASIR-V) techniques at different scan doses using a phantom with liver density.

Methods

The Gammex computed tomography (CT) phantom with a standard liver-density insert (ρew=1.06) was scanned at six different radiation doses (CTDIvol): 30, 20, 15, 10, 7.5, and 4.5 mGy). Images obtained at each dose were reconstructed using DLIR and ASIR-V. Image quality was analyzed through the imQuest software. The quality of reconstructed images by DLIR at 4.5 mGy (lowest radiation dose) and ASIR-V at 15 mGy (recommended scan dose) were compared using the Bland–Altman method.

Results

Across the six doses, DLIR significantly outperformed ASIR-V in key metrics, such as noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and detectability index (\begin{document}$d' $\end{document}). Bland–Altman analysis indicated that the quality of reconstructed images by DLIR at 4.5 mGy was significantly better to those by ASIR-V at 15 mGy. The noise level of DLIR images at 4.5 mGy was (17.41±0.32), which is significantly lower than that of ASIR-V at 15 mGy (21.17±0.67). At 4.5 mGy, DLIR SNR, CNR, and d’ were (3.21±0.24), (3.42±0.35), and (8.81±0.63), respectively, which are significantly higher than that of ASIR-V at 15 mGy (2.69±0.14), (2.87±0.11), and (5.61±1.28), respectively.

Conclusion

In CT scan of focal liver-density lesions using a phantom, DLIR significantly improved the SNR, CNR, and d’ values and reduced image noise compared to ASIR-V. DLIR was able to achieve better quality image reconstruction at 4.5 mGy than the conventional ASIR-V reconstruction at 15 mGy.

Schlagwörter

deep learningphantomctradiation dose

Zitieren

APA

PAN, Z., LIU, L., LI, Q., QU, T., ZHANG, S., & XIE, X. (2025). Deep Learning Image Reconstruction to Improve Computed Tomography Image Quality of the Phantom with Standard Liver Density. CT Lilun yu yingyong yanjiu, 34(4), 677-685. https://doi.org/10.15953/j.ctta.2024.056

Vancouver

PAN Z, LIU L, LI Q, QU T, ZHANG S, XIE X. Deep Learning Image Reconstruction to Improve Computed Tomography Image Quality of the Phantom with Standard Liver Density. CT Lilun yu yingyong yanjiu. 2025;34(4):677-685. doi:10.15953/j.ctta.2024.056

BibTeX
@article{PAN2025,
  author = {PAN, Zhijie and LIU, Ling and LI, Qingyao and QU, Tingting and ZHANG, Shuai and XIE, Xueqian},
  title = {Deep Learning Image Reconstruction to Improve Computed Tomography Image Quality of the Phantom with Standard Liver Density},
  journal = {CT Lilun yu yingyong yanjiu},
  year = {2025},
  volume = {34},
  number = {4},
  pages = {677--685},
  doi = {10.15953/j.ctta.2024.056}
}
Kostenlos starten →