EUVIMEDEUVIMED entdecken
Open Access

Infections Associated with Medtronic Duet External Ventricular Drains — Rhode Island Hospital, Providence, Rhode Island, January 2023–January 2024

Kevin M Gibas, Dianne Auld, Stephanie Parente, Jean Horoho, Leonard A Mermel

MMWR. Morbidity and Mortality Weekly Report · 11.04.2024 · 73 (14) · 312-316

DOI: 10.15585/mmwr.mm7314a4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift MMWR. Morbidity and Mortality Weekly Report, online erschienen am 11.04.2024. Verfasst von Kevin M Gibas und 4 weiteren Autor:innen. Hauptthemen laut MEDLINE: Cross Infection. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

External ventricular drains (EVDs) are medical devices that are inserted into the ventricles of the brain to drain excess fluid, manage intracranial hypertension, monitor intracranial pressure, and administer medications. Unintentional disconnections and breaks or fractures (breaks) of EVDs or associated drainage system components can result in cerebrospinal fluid (CSF) leakage and increased risk for EVD-associated infections. After replacement of Integra Life Sciences EVD systems with Medtronic Duet EVD systems at Rhode Island Hospital in mid-September 2023, a threefold increase was observed in the prevalence of positive CSF cultures, from 2.8 per 1,000 days with an EVD in place (EVD days) during January-September 2023 to 11.4 per 1,000 EVD days during October 2023-January 2024 (rate ratio [RR] = 5.7; 95% CI = 1.5-22.0; p = 0.01) and an eightfold increase in the prevalence of infections, from 0.7 to 6.5 per 1,000 EVD days (RR = 9.8; 95% CI = 1.1-87.3; p = 0.04). An investigation by Rhode Island Hospital Infection Control during December 2023-January 2024 identified frequent reports of disconnections and breaks of the Medtronic Duet EVD system. A search of the Food and Drug Administration Manufacturer and User Facility Device Experience database identified 326 reports nationwide of disconnection and breaks of components of the Duet EVD system, including 175 during 2023. A Medical Product Safety Network report was filed. The Duet EVD product was ultimately recalled in January 2024, citing disconnections of the EVD system and reports of CSF leakage and infection. Given the widespread use of EVD systems by neurosurgery centers and the risk for EVD-associated infections, a strategy for future consideration by hospital infection prevention and control programs might be inclusion of EVD-associated infections in hospital surveillance programs to rapidly identify increases in these events and determine factors related to such infections to prevent additional infections.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cross InfectionUnited StatesHumansRhode IslandDrainageBrainHospitalsRetrospective Studies

Zitieren

APA

Gibas, K. M., Auld, D., Parente, S., Horoho, J., & Mermel, L. A. (2024). Infections Associated with Medtronic Duet External Ventricular Drains — Rhode Island Hospital, Providence, Rhode Island, January 2023–January 2024. MMWR. Morbidity and Mortality Weekly Report, 73(14), 312-316. https://doi.org/10.15585/mmwr.mm7314a4

Vancouver

Gibas KM, Auld D, Parente S, Horoho J, Mermel LA. Infections Associated with Medtronic Duet External Ventricular Drains — Rhode Island Hospital, Providence, Rhode Island, January 2023–January 2024. MMWR. Morbidity and Mortality Weekly Report. 2024;73(14):312-316. doi:10.15585/mmwr.mm7314a4

BibTeX
@article{Gibas2024,
  author = {Gibas, Kevin M and Auld, Dianne and Parente, Stephanie and Horoho, Jean and Mermel, Leonard A},
  title = {Infections Associated with Medtronic Duet External Ventricular Drains — Rhode Island Hospital, Providence, Rhode Island, January 2023–January 2024},
  journal = {MMWR. Morbidity and Mortality Weekly Report},
  year = {2024},
  volume = {73},
  number = {14},
  pages = {312--316},
  doi = {10.15585/mmwr.mm7314a4}
}
Kostenlos starten →