Vergleichsstudie in der Zeitschrift Physiology International, 2025. Verfasst von Dániel Kulin und 6 weiteren Autor:innen. Beteiligte Einrichtung: 1Institute of Translational Medicine - Semmelweis University, Budapest, Hungary. Hauptthemen laut MEDLINE: Photoplethysmography, Echocardiography, Pulse Wave Analysis, Heart. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
This study assesses the utility of photoplethysmography (PPG) as a non-invasive method to evaluate cardiac function, addressing the critical need for accessible biomarkers in various cardiovascular conditions, including heart failure management.
Methods
By conducting simultaneous echocardiography and PPG measurements on 37 healthy volunteers, we analyzed both traditional and novel composite pulse wave scores to correlate peripheral PPG data with central echocardiographic outcomes.
Results
Our results show a good correlation between PPG-based and echocardiography-derived ejection times ( r = 0.648, P < 0.001), though Bland-Altmann analysis results reveal that PPG consistently overestimated ejection times by a mean difference of +95 ms. Moreover, eleven PPG parameters significantly correlated with key echocardiographic indicators of systolic and diastolic function, such as left ventricular dimensions, global longitudinal strain, aortic functionality, atrial contraction (MV-A), and ventricular filling pressure (E/e’ lat) with clinical relevance indicated by correlations ( r ) above 0.4 ( P < 0.05).
Conclusion
The findings pave the way for further studies in various patient groups to explore the potential of PPG in enhancing home monitoring and regular cardiovascular assessments. This work not only broadens our understanding of the physiological relationships between peripheral and central cardiovascular measures but also introduces innovative metrics that might bring some added value to the current standards of patient care by facilitating early detection and personalized management of heart conditions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 1Institute of Translational Medicine - Semmelweis University, Budapest, Hungary.
- 2E-Med4All Europe Kft, Budapest, Hungary.
- 3Heart and Vascular Centre - Semmelweis University, Budapest, Hungary.
- 4National Korányi Institute for Pulmonology, Budapest, Hungary.
ORCID-Profile: Dániel Kulin, Zsuzsanna Miklós
Zitieren
APA
Kulin, D., Antali, F., Horváth, M., Kulin, S., Kulin, S., Miklós, Z., & Szűcs, A. (2025). Evaluating photoplethysmography-based pulsewave parameters and composite scores for assessment of cardiac function: A comparison with echocardiography. Physiology International. https://doi.org/10.1556/2060.2025.00675
Vancouver
Kulin D, Antali F, Horváth M, Kulin S, Kulin S, Miklós Z, et al. Evaluating photoplethysmography-based pulsewave parameters and composite scores for assessment of cardiac function: A comparison with echocardiography. Physiology International. 2025. doi:10.1556/2060.2025.00675
BibTeX
@article{Kulin2025,
author = {Kulin, Dániel and Antali, Flóra and Horváth, Márton and Kulin, Sándor and Kulin, Sándor and Miklós, Zsuzsanna and Szűcs, Andrea},
title = {Evaluating photoplethysmography-based pulsewave parameters and composite scores for assessment of cardiac function: A comparison with echocardiography},
journal = {Physiology International},
year = {2025},
doi = {10.1556/2060.2025.00675}
}
EUVIMED