Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel

Karim Ghazikhanlou-Sani, Azizollah Rahimi, Maryam Poorkaveh, Samira Eynali, Fereshteh Koosha, Mohsen Shoja

Interventional Medicine and Applied Science · 2018

DOI: 10.1556/1646.10.2018.18 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Interventional Medicine and Applied Science, 2018. Verfasst von Karim Ghazikhanlou-Sani und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Paramedical School, Hamadan University of Medical Sciences, Hamadan, Iran. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Operating rooms in hospitals are facilitated with different types of electronic systems, which produce electromagnetic waves. High intensities of magnetic waves may have harmful effects on biological environments. This study aims to evaluate the electromagnetic field intensity at different parts of operating rooms at the first stage and estimate the occupational exposure to operating room personnel at the next phase. Materials and methods At this cross-sectional study, the magnetic field intensity was evaluated using teslameter at several parts of operating rooms, during operating procedures, while electrical instruments were working. Background electromagnetic field intensity was measured when all the electrical systems were idle. Statistical analysis was performed using SPSS software. The results were compared with ICNIRP standards.

Results

The maximum intensity of magnetic field was measured around high-voltage systems at the distance of 50 cm in the personnel’s standing area at DCR and PCNL operating procedures were 5.9 and 5.6, respectively. The number of on-mode electrical systems was inconsistent with the intensity of electromagnetic fields at the standing area of operating room personnel’s. The intensity of magnetic fields around high-voltage systems, which was about 46.75 mG at the distance of 10 cm, was the highest among measured electromagnetic fields.

Conclusions

The highest magnetic field intensity measured in this study was related to high-voltage systems and is lower than advised intensity by ICNIRP for occupational exposure. Based on this study, it can be concluded that there are no considerable risks of electromagnetic exposure for operating room personnels.

Schlagwörter

Stichwörter der Autor:innen

electromagnetic wavesintensitymagnetic fieldoccupational exposureoperating room

Autor:innen und Institutionen

  • Department of Radiology, Paramedical School, Hamadan University of Medical Sciences, Hamadan, Iran.
  • Faculty of Medicine, Department of Medical Physics, Ahvaz Judishapur University of Medical Sciences, Ahvaz, Iran.
  • Radiation Biology Research Center, Iran University of Medical Sciences, Tehran, Iran.
  • Faculty of Medicine, Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran.
  • Department of Radiology, School of Allied Medical Sciences, Semnan University of Medical Sciences, Semnan, Iran.

Zitieren

APA

Ghazikhanlou-Sani, K., Rahimi, A., Poorkaveh, M., Eynali, S., Koosha, F., & Shoja, M. (2018). Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel. Interventional Medicine and Applied Science. https://doi.org/10.1556/1646.10.2018.18

Vancouver

Ghazikhanlou-Sani K, Rahimi A, Poorkaveh M, Eynali S, Koosha F, Shoja M. Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel. Interventional Medicine and Applied Science. 2018. doi:10.1556/1646.10.2018.18

BibTeX
@article{GhazikhanlouSani2018,
  author = {Ghazikhanlou-Sani, Karim and Rahimi, Azizollah and Poorkaveh, Maryam and Eynali, Samira and Koosha, Fereshteh and Shoja, Mohsen},
  title = {Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel},
  journal = {Interventional Medicine and Applied Science},
  year = {2018},
  doi = {10.1556/1646.10.2018.18}
}
Kostenlos starten →